» Articles » PMID: 3559812

Dynamic Trend Monitoring of Cerebral Blood Flow Velocity in Newborn Infants

Overview
Journal J Pediatr
Specialty Pediatrics
Date 1987 Apr 1
PMID 3559812
Citations 2
Authors
Affiliations
Soon will be listed here.
Abstract

A method for obtaining reproducible Doppler cerebral blood flow velocity (CBFV) measurements over several hours, utilizing a probe fixation technique and an on-line system for making beat-to-beat area under the curve (AUC) measurements, is described. Thirty heartbeat samples were collected at three time intervals over 3 hours in a group of 22 healthy preterm infants. The stability of the AUC measurements despite changes in behavioral state was demonstrated by randomized block ANOVA. The sampling error was +/- 6.92%, with 95% confidence intervals of +/- 13.1%. By facilitating reproducible serial measurements of CBFV, this trend monitoring technique encourages a dynamic view of cerebrovascular control, and may be used to probe central nervous system autoregulatory pathophysiology and to evaluate the effects of therapeutic interventions.

Citing Articles

NeoDoppler: New ultrasound technology for continous cerebral circulation monitoring in neonates.

Vik S, Torp H, Follestad T, Stoen R, Nyrnes S Pediatr Res. 2019; 87(1):95-103.

PMID: 31404920 PMC: 6960092. DOI: 10.1038/s41390-019-0535-0.


Acoustic power measurements of Doppler ultrasound devices used for perinatal and infant examinations.

Rabe H, Grohs B, SCHMIDT R, Schloo R, Bomelburg T, Jorch G Pediatr Radiol. 1990; 20(4):277-81.

PMID: 2186347 DOI: 10.1007/BF02019666.