Ultrafast 30-s Total-body PET/CT Scan: a Preliminary Study

Overview
Date 2022 May 16
PMID 35578037
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: The aim of this study is to explore the diagnostic value of the images obtained in ultrafast 30-s acquisition time by the total-body PET/CT (F-FDG injection dose of about 3.7 MBq/kg), and to evaluate whether they can meet the requirements of clinical diagnosis or not.

Methods: This retrospective study explored the clinical value of ultrafast 30-s F-FDG total-body PET/CT in 88 oncology patients, using the post-surgical pathological diagnosis as the reference standard. The data were acquired over 300 s and reconstructed using all 300-s data (G300) and only the initial 30 s (G30). Two readers independently assessed all images qualitatively and quantitatively. The diagnostic performance was compared between G300 and G30.

Results: The G300 average qualitative score was higher than G30 (P < 0.001). G300 and G30 also differed quantitatively in the liver and mediastinum SUV, SD, and SNR (all P < 0.001), but had similar sensitivities (89.09% vs. 86.36%, P = 0.250). The G300 group had higher accuracy (79.73%) and a larger area under the curve (0.709) than G30 (77.70% and 0.695, respectively; all P > 0.05).

Conclusion: The 30-s total-body PET/CT could meet clinical diagnostic requirements for malignant tumors in patients intolerant to prolonged horizontal positioning.

Citing Articles

Leveraging small voxel with optimal acquisition time for [F]mFBG total-body PET/CT imaging in pediatric patients with neuroblastoma: a preliminary study.

Cheng Z, Deng X, Song S, Wu Y, Tang H, Zou S Eur J Nucl Med Mol Imaging. 2025; .

PMID: 39888420 DOI: 10.1007/s00259-025-07098-1.


The SwiftScan step-and-shoot continuous mode improves SPECT scanning efficiency: a preliminary phantom and clinical test.

Li J, Zhang K, Pang X, Huang L, Tian X, Liu J EJNMMI Phys. 2025; 12(1):1.

PMID: 39745654 PMC: 11695528. DOI: 10.1186/s40658-024-00709-0.


Expert consensus on workflow of PET/CT with long axial field-of-view.

Liu G, Gu Y, Sollini M, Lazar A, Besson F, Li S Eur J Nucl Med Mol Imaging. 2024; 52(3):1038-1049.

PMID: 39520515 DOI: 10.1007/s00259-024-06968-4.


Sedation-free pediatric [F]FDG imaging on totalbody PET/CT with the assistance of artificial intelligence.

Zhou X, Xue S, Li L, Seifert R, Dong S, Chen R Eur J Nucl Med Mol Imaging. 2024; 51(13):4062-4072.

PMID: 38958680 DOI: 10.1007/s00259-024-06818-3.


Intelligent ultrafast total-body PET for sedation-free pediatric [F]FDG imaging.

Zhou X, Fu Y, Dong S, Li L, Xue S, Chen R Eur J Nucl Med Mol Imaging. 2024; 51(8):2353-2366.

PMID: 38383744 DOI: 10.1007/s00259-024-06649-2.


References
1.
Delbeke D, Coleman R, Guiberteau M, Brown M, Royal H, Siegel B . Procedure guideline for tumor imaging with 18F-FDG PET/CT 1.0. J Nucl Med. 2006; 47(5):885-95. View

2.
Boellaard R, Krak N, Hoekstra O, Lammertsma A . Effects of noise, image resolution, and ROI definition on the accuracy of standard uptake values: a simulation study. J Nucl Med. 2004; 45(9):1519-27. View