Calculation of Attenuation Parameter for Ir-192 Gamma Source in Shielding Materials

Overview
Journal J Med Phys
Date 2022 May 13
PMID 35548040
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: Calculation of photon attenuation is necessary for the selection of shielding materials for an irradiation facility.

Methods And Materials: In this work, a Monte Carlo simulation was utilized to assess the effectiveness of clay-polyethylene mixture and clay as the radiation shielding materials for high-energy gamma sources (Ir-192). Ordinary concrete was also studied as the benchmark.

Results: The calculated linear attenuation values for ordinary concrete are within 0.44% of the standard XCOM value for 380 keV photon. In the case of a multienergy Ir-192 gamma source, the calculated linear attenuation coefficient (μ) for ordinary concrete is 15.5% and 7.25% higher than clay and fabricated clay-polyethylene, respectively. Meanwhile, the μ value for fabricated clay-polyethylene is 8.3% higher than that of clay.

Conclusion: In conclusion, a 10 cm thickness of clay and clay-polyethylene mixture is sufficient to attenuate 87% and 89% of photons from Ir-192 source. The calculated linear attenuation coefficients for the three shielding materials are also consistently higher, about 7.5%, than that of the XCOM value for 380 keV photon.

Citing Articles

Photon dose rate distribution inside and outside a brachytherapy room.

Aminordin Sabri A, Tajudin S, Abdul Aziz M, Furuta T Radiol Phys Technol. 2023; 16(1):109-117.

PMID: 36729272 DOI: 10.1007/s12194-023-00703-8.

References
1.
Wani A, Ara A, Usmani J . Lead toxicity: a review. Interdiscip Toxicol. 2016; 8(2):55-64. PMC: 4961898. DOI: 10.1515/intox-2015-0009. View

2.
McCaffrey J, Mainegra-Hing E, Shen H . Optimizing non-Pb radiation shielding materials using bilayers. Med Phys. 2010; 36(12):5586-94. DOI: 10.1118/1.3260839. View

3.
Hyun S, Kim K, Jahng T, Kim H . Efficiency of lead aprons in blocking radiation - how protective are they?. Heliyon. 2016; 2(5):e00117. PMC: 4946218. DOI: 10.1016/j.heliyon.2016.e00117. View

4.
Yue K, Luo W, Dong X, Wang C, Wu G, Jiang M . A new lead-free radiation shielding material for radiotherapy. Radiat Prot Dosimetry. 2009; 133(4):256-60. DOI: 10.1093/rpd/ncp053. View

5.
Flora G, Gupta D, Tiwari A . Toxicity of lead: A review with recent updates. Interdiscip Toxicol. 2012; 5(2):47-58. PMC: 3485653. DOI: 10.2478/v10102-012-0009-2. View