Evaluation of Radiation Dose Reduction and Its Effect on Image Quality for Different Flat-Panel Detectors

Overview
Journal J Med Phys
Date 2022 May 13
PMID 35548029
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To evaluate the image quality of semi-anatomical chest radiographs acquired using low radiation doses from seven different flat-panel detector (FPD) systems.

Materials And Methods: Radiographs of a semi-anatomical chest phantom were acquired at 70 and 110 kVp using 7 different FPDs from 5 vendors. Radiation doses were measured using a dose-area-product (DAP) meter. To standardize measurements across all FPDs, DAP value of 51.05 μGym obtained at 70 kVp and 9.43 μGym at 110 kVp was used as reference in this study. Radiation doses were reduced by manually adjusting mAs for both tube potentials in all FPD systems to achieve acceptable image quality. Contrast-to-noise ratio, signal-to-noise ratio and figure of merit (FOM) in lung, heart, and diaphragm for all images were analyzed.

Results: In comparison with set standard protocol, radiation dose reduction of 20%, 21%, 21.6%, 59.5%, 60.7%, 62.2%, and 67.6% with optimal image quality was observed in Prognosys Prorad, GE Definium 8000, Siemens Fusion, Fujifilm FGX, Fujifilm FGXR, Philips Digital Diagnost and Siemens Aristos at 70 kVp. At 110 kVp, dose reduction of 15.7% and 34.8% was possible only for Philips Digital Diagnost and Siemens Aristos. FOM was high at 110 kVp even when radiation doses were reduced by a factor 2 when compared to 70 kVp in all digital radiography systems.

Conclusion: This study demonstrates the feasibility of using a semi-anatomical chest phantom in the optimization of radiation dose and image quality. The FOM was a good indicator in assessing image quality between different detectors.

Citing Articles

Development and clinical feasibility of a reduced-dose radiograph in children for feeding tube placement.

Kaplan S, Jalloul M, Akbari E, White A, Shumyatsky G, Flowers C Pediatr Radiol. 2023; 54(2):218-227.

PMID: 38141080 DOI: 10.1007/s00247-023-05829-w.

References
1.
Willis C, Slovis T . The ALARA concept in pediatric CR and DR: dose reduction in pediatric radiographic exams--a white paper conference executive summary. Pediatr Radiol. 2004; 34 Suppl 3:S162-4. DOI: 10.1007/s00247-004-1264-y. View

2.
Butler M, Rainford L, Last J, Brennan P . Are exposure index values consistent in clinical practice? A multi-manufacturer investigation. Radiat Prot Dosimetry. 2010; 139(1-3):371-4. DOI: 10.1093/rpd/ncq094. View

3.
Rong X, Shaw C, Liu X, Lemacks M, Thompson S . Comparison of an amorphous silicon/cesium iodide flat-panel digital chest radiography system with screen/film and computed radiography systems--a contrast-detail phantom study. Med Phys. 2002; 28(11):2328-35. DOI: 10.1118/1.1408620. View

4.
Samei E, Flynn M . An experimental comparison of detector performance for direct and indirect digital radiography systems. Med Phys. 2003; 30(4):608-22. DOI: 10.1118/1.1561285. View

5.
Chotas H, Floyd Jr C, Johnson G, Ravin C . Quality control phantom for digital chest radiography. Radiology. 1997; 202(1):111-6. DOI: 10.1148/radiology.202.1.8988199. View