Robustness Assessment of Images From a 0.35T Scanner of an Integrated MRI-Linac: Characterization of Radiomics Features in Phantom and Patient Data

Overview
Authors
Affiliations
Soon will be listed here.
Abstract

Radiomics entails the extraction of quantitative imaging biomarkers (or radiomics features) hypothesized to provide additional pathophysiological and/or clinical information compared to qualitative visual observation and interpretation. This retrospective study explores the variability of radiomics features extracted from images acquired with the 0.35 T scanner of an integrated MRI-Linac. We hypothesized we would be able to identify features with high repeatability and reproducibility over various imaging conditions using phantom and patient imaging studies. We also compared findings from the literature relevant to our results. Eleven scans of a Magphan RT phantom over 13 months and 11 scans of a ViewRay Daily QA phantom over 11 days constituted the phantom data. Patient datasets included 50 images from ten anonymized stereotactic body radiation therapy (SBRT) pancreatic cancer patients (50 Gy in 5 fractions). A True Fast Imaging with Steady-State Free Precession (TRUFI) pulse sequence was selected, using a voxel resolution of 1.5 mm × 1.5 mm × 1.5 mm and 1.5 mm × 1.5 mm × 3.0 mm for phantom and patient data, respectively. A total of 1087 shape-based, first, second, and higher order features were extracted followed by robustness analysis. Robustness was assessed with the Coefficient of Variation (CoV < 5%). We identified 130 robust features across the datasets. Robust features were found within each category, except for 2 second-order sub-groups, namely, Gray Level Size Zone Matrix (GLSZM) and Neighborhood Gray Tone Difference Matrix (NGTDM). Additionally, several robust features agreed with findings from other stability assessments or predictive performance studies in the literature. We verified the stability of the 0.35 T scanner of an integrated MRI-Linac for longitudinal radiomics phantom studies and identified robust features over various imaging conditions. We conclude that phantom measurements can be used to identify robust radiomics features. More stability assessment research is warranted.

Citing Articles

Radiogenomics Pilot Study: Association Between Radiomics and Single Nucleotide Polymorphism-Based Microarray Copy Number Variation in Diagnosing Renal Oncocytoma and Chromophobe Renal Cell Carcinoma.

Alhussaini A, Veluchamy A, Jawli A, Kernohan N, Tang B, Palmer C Int J Mol Sci. 2024; 25(23).

PMID: 39684226 PMC: 11641663. DOI: 10.3390/ijms252312512.


Magnetic Resonance-Guided Cancer Therapy Radiomics and Machine Learning Models for Response Prediction.

Fajemisin J, Gonzalez G, Rosenberg S, Ullah G, Redler G, Latifi K Tomography. 2024; 10(9):1439-1454.

PMID: 39330753 PMC: 11435563. DOI: 10.3390/tomography10090107.


Robustness of radiomics features on 0.35 T magnetic resonance imaging for magnetic resonance-guided radiotherapy.

Michalet M, Valenzuela G, Debuire P, Riou O, Azria D, Nougaret S Phys Imaging Radiat Oncol. 2024; 31:100613.

PMID: 39140002 PMC: 11320460. DOI: 10.1016/j.phro.2024.100613.


Advancements in MRI-Based Radiomics and Artificial Intelligence for Prostate Cancer: A Comprehensive Review and Future Prospects.

Chaddad A, Tan G, Liang X, Hassan L, Rathore S, Desrosiers C Cancers (Basel). 2023; 15(15).

PMID: 37568655 PMC: 10416937. DOI: 10.3390/cancers15153839.


An integrated and fast imaging quality assurance phantom for a 0.35 T magnetic resonance imaging linear accelerator.

Sohn J, Lim S, Das I, Yadav P Phys Imaging Radiat Oncol. 2023; 27:100462.

PMID: 37449023 PMC: 10338140. DOI: 10.1016/j.phro.2023.100462.

References
1.
Gillies R, Kinahan P, Hricak H . Radiomics: Images Are More than Pictures, They Are Data. Radiology. 2015; 278(2):563-77. PMC: 4734157. DOI: 10.1148/radiol.2015151169. View

2.
Collewet G, Strzelecki M, Mariette F . Influence of MRI acquisition protocols and image intensity normalization methods on texture classification. Magn Reson Imaging. 2004; 22(1):81-91. DOI: 10.1016/j.mri.2003.09.001. View

3.
Mittauer K, Dunkerley D, Yadav P, Bayouth J . Characterization and longitudinal assessment of daily quality assurance for an MR-guided radiotherapy (MRgRT) linac. J Appl Clin Med Phys. 2019; 20(11):27-36. PMC: 6839363. DOI: 10.1002/acm2.12735. View

4.
Saenz D, Yan Y, Christensen N, Henzler M, Forrest L, Bayouth J . Characterization of a 0.35T MR system for phantom image quality stability and in vivo assessment of motion quantification. J Appl Clin Med Phys. 2015; 16(6):30-40. PMC: 5691014. DOI: 10.1120/jacmp.v16i6.5353. View

5.
Wong O, Yuan J, Zhou Y, Yu S, Cheung K . Longitudinal acquisition repeatability of MRI radiomics features: An ACR MRI phantom study on two MRI scanners using a 3D T1W TSE sequence. Med Phys. 2020; 48(3):1239-1249. DOI: 10.1002/mp.14686. View