» Articles » PMID: 35516795

Clinical Utility of HeartLogic, a Multiparametric Telemonitoring System, in Heart Failure

Abstract

Telemonitoring through multiple variables measured on cardiac devices has the potential to improve the follow-up of patients with heart failure. The HeartLogic algorithm (Boston Scientific), implemented in some implantable cardiac defibrillators and cardiac resynchronisation therapy, allows monitoring of the nocturnal heart rate, respiratory movements, thoracic impedance, physical activity and the intensity of heart tones, with the aim of predicting major clinical events. Although HeartLogic has demonstrated high sensitivity for the detection of heart failure decompensations, its effects on hospitalisation and mortality in randomised clinical trials has not yet been corroborated. This review details how the HeartLogic algorithm works, compiles available evidence from clinical studies, and discusses its application in daily clinical practice.

Citing Articles

Prediction of heart failure events based on physiologic sensor data in HINODE defibrillator patients.

Nishii N, Sakata Y, Murohara T, Ando K, Ikeda T, Mitsuhashi T ESC Heart Fail. 2024; 11(5):3322-3331.

PMID: 38956896 PMC: 11424318. DOI: 10.1002/ehf2.14890.


Clinical Predictive Modeling of Heart Failure: Domain Description, Models' Characteristics and Literature Review.

Odrobina I Diagnostics (Basel). 2024; 14(4).

PMID: 38396482 PMC: 10888082. DOI: 10.3390/diagnostics14040443.


The patient perspective on remote monitoring of implantable cardiac devices.

Hillmann H, Hansen C, Przibille O, Duncker D Front Cardiovasc Med. 2023; 10:1123848.

PMID: 36937908 PMC: 10017432. DOI: 10.3389/fcvm.2023.1123848.


Personalized Therapy and Clinical Outcome for Heart Failure.

Acena A, de Juan Baguda J, Rincon L J Clin Med. 2022; 11(16).

PMID: 36013089 PMC: 9410394. DOI: 10.3390/jcm11164851.

References
1.
Whellan D, Ousdigian K, Al-Khatib S, Pu W, Sarkar S, Porter C . Combined heart failure device diagnostics identify patients at higher risk of subsequent heart failure hospitalizations: results from PARTNERS HF (Program to Access and Review Trending Information and Evaluate Correlation to Symptoms in Patients.... J Am Coll Cardiol. 2010; 55(17):1803-10. DOI: 10.1016/j.jacc.2009.11.089. View

2.
Auricchio A, Gold M, Brugada J, Nolker G, Arunasalam S, Leclercq C . Long-term effectiveness of the combined minute ventilation and patient activity sensors as predictor of heart failure events in patients treated with cardiac resynchronization therapy: Results of the Clinical Evaluation of the Physiological.... Eur J Heart Fail. 2014; 16(6):663-70. DOI: 10.1002/ejhf.79. View

3.
Givertz M, Stevenson L, Costanzo M, Bourge R, Bauman J, Ginn G . Pulmonary Artery Pressure-Guided Management of Patients With Heart Failure and Reduced Ejection Fraction. J Am Coll Cardiol. 2017; 70(15):1875-1886. DOI: 10.1016/j.jacc.2017.08.010. View

4.
Guedon-Moreau L, Lacroix D, Sadoul N, Clementy J, Kouakam C, Hermida J . A randomized study of remote follow-up of implantable cardioverter defibrillators: safety and efficacy report of the ECOST trial. Eur Heart J. 2012; 34(8):605-14. PMC: 3578267. DOI: 10.1093/eurheartj/ehs425. View

5.
Gardner R, Thakur P, Hammill E, Nair D, Eldadah Z, Stancak B . Multiparameter diagnostic sensor measurements during clinically stable periods and worsening heart failure in ambulatory patients. ESC Heart Fail. 2021; 8(2):1571-1581. PMC: 8006698. DOI: 10.1002/ehf2.13261. View