» Articles » PMID: 35377531

Reconstruction of Paprosky Type III Acetabular Defects by Three-Dimensional Printed Porous Augment: Techniques and Clinical Outcomes of 18 Consecutive Cases

Overview
Journal Orthop Surg
Specialty Orthopedics
Date 2022 Apr 4
PMID 35377531
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To introduce the surgical technique of reconstruction of Paprosky type III acetabular defects by 3D printed porous augments.

Methods: First, CT scans of pelvis were obtained to establish the 3D reconstruction model of 3D printed porous augment. Then, a nylon pelvis model was printed to simulate operation with the surgeons. At this time, the augment was designed and modified according to the surgeon's suggestions and the 3D printing principles. Eighteen patients with Paprosky type III acetabular defects receiving reconstructive surgery by 3D printed porous augments were included in current study. Their data, including general information, intra-operative findings, imaging results, functional scores, and complications were retrospectively analyzed.

Results: The mean follow-up time lasted 33.3 ± 2.0 (24-56) months. The average limb-length discrepancy (LLD) was 31.7 ± 4.2 (3-59) mm preoperatively, 7.7 ± 1.4 (1-21) mm postoperatively (P < 0.0001), and 7.5 ± 1.2 (0-18) mm at the latest follow-up. The mean vertical position of hip center of rotation (HCOR) from the interteardrop line changed from preoperative 50.7 ± 3.9 (23.3-75.3) mm to postoperative 22.9 ± 1.9 (10.1-40.3) mm (P < 0.0001), with the latest follow-up revealing an HCOR of 22.3 ± 1.7 (11.0-40.5) mm. Follow-up study showed that no hip had radiolucencies and radiological loosening of the acetabular components and augment. The average Harris hip score (HHS) improved from 40.3 ± 4.5 (10.5-71) before operation to 88.4 ± 1.9 (75-97) at the last follow-up (P < 0.0001). Moreover, follow-up exhibited that no periprosthetic joint infection, hip dislocation, fracture, and re-revision occurred.

Conclusion: Surgical treatment of Paprosky type III acetabular defect with 3D printed porous augment was simple, achieved good match between porous augment and the defect bone surface and the acetabular component, ideally restored LLD and HCOR after operation, significantly improved HHS and attained good early clinical outcomes. It is a promising personalized solution for patients with severe acetabular bone defect.

Citing Articles

Three dimensionalprinted titanium block to reconstruct severe acetabular bone defects in primary hip arthroplasty.

Sun H, Tan R, Liu J, Shao S, Xiang J, Li S Int Orthop. 2025; 49(3):595-603.

PMID: 39921749 DOI: 10.1007/s00264-025-06444-0.


[Research progress of three-dimensional printed customized prosthesis and its application in acetabular reconstruction of hip revision surgery].

Zhang H, Ma X, Li B, Li K, Liu Y, Zhou J Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2024; 38(11):1414-1420.

PMID: 39542636 PMC: 11563742. DOI: 10.7507/1002-1892.202406002.


Application of Intraoperative Individualized Cup Combination-Cup-on-Cup Technique in Severe Acetabular Defects.

Li W, Liu D, Liu X, Li X Orthop Surg. 2024; 16(11):2853-2861.

PMID: 39311082 PMC: 11541136. DOI: 10.1111/os.14248.


Integrated Acetabular Prosthesis Versus Bone Grafting in Total Hip Arthroplasty for Crowe Type II and III Hip Dysplasia: A Retrospective Case-Control Study.

Cheng L, Liu Y, Wang L, Ying J, Li J, Wang F Orthop Surg. 2024; 16(10):2401-2409.

PMID: 39043609 PMC: 11456704. DOI: 10.1111/os.14143.


Standardized 3D-printed trabecular titanium augment and cup for acetabular bone defects in revision hip arthroplasty: a mid-term follow-up study.

Hao L, Zhang Y, Bian W, Song W, Li K, Wang N J Orthop Surg Res. 2023; 18(1):521.

PMID: 37481549 PMC: 10362760. DOI: 10.1186/s13018-023-03986-0.


References
1.
Lal H, Patralekh M . 3D printing and its applications in orthopaedic trauma: A technological marvel. J Clin Orthop Trauma. 2018; 9(3):260-268. PMC: 6128305. DOI: 10.1016/j.jcot.2018.07.022. View

2.
Van Kleunen J, Lee G, Lementowski P, Nelson C, Garino J . Acetabular revisions using trabecular metal cups and augments. J Arthroplasty. 2009; 24(6 Suppl):64-8. DOI: 10.1016/j.arth.2009.02.001. View

3.
DeLee J, CHARNLEY J . Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976; (121):20-32. View

4.
Schreurs B, Luttjeboer J, Thien T, de Waal Malefijt M, Buma P, Veth R . Acetabular revision with impacted morselized cancellous bone graft and a cemented cup in patients with rheumatoid arthritis. A concise follow-up, at eight to nineteen years, of a previous report. J Bone Joint Surg Am. 2009; 91(3):646-51. DOI: 10.2106/JBJS.G.01701. View

5.
Banerjee S, Issa K, Kapadia B, Pivec R, Khanuja H, Mont M . Systematic review on outcomes of acetabular revisions with highly-porous metals. Int Orthop. 2013; 38(4):689-702. PMC: 3971260. DOI: 10.1007/s00264-013-2145-5. View