Evaluation of Corvis ST Tonometer with the Updated Software in Glaucoma Practice

Overview
Specialty Ophthalmology
Date 2022 Mar 21
PMID 35310063
Authors
Affiliations
Soon will be listed here.
Abstract

Aim: To evaluate the agreement of biomechanically corrected intraocular pressure (b-IOP) and central corneal thickness (CCT) measurements obtained with the updated Corvis ST tonometer versus Goldmann applanation tonometry (GAT) and optical-based corneal pachymetry (OB-CCT) in controls, patients with ocular hypertension (OHT) and primary open angle glaucoma (POAG). Additionally, we examined the differences in corneal deformation parameters provided by the updated Corvis ST among the three groups.

Methods: For each participant, GAT IOP, OB-CCT and measurements with a Corvis ST with updated software were obtained. Bland-Altman analysis was used to assess the agreement between the two measurement methods.

Results: A consecutive series of 80 eyes from 80 participants (30 with POAG, 25 with OHT and 25 normal controls) were included in this prospective study. The mean GAT IOP of all eyes was 17.2±3.6 mm Hg, and the mean b-IOP was 15.9±3.7 mm Hg (Spearman's rho=0.767, <0.001). The 95% limits of agreement (LoAs) ranged from -3.1 mm Hg to 5.5 mm Hg for GAT IOP and b-IOP. b-IOP was not correlated with OB-CCT (Spearman's rho=-0.13 =0.917). Meanwhile there was a weak positive corelation between OB-CCT and GAT IOP-b-IOP difference (Spearman's rho=0.378, =0.001). The mean OB-CCT was 549.5±36.4 µm, and the Corvis-CCT was 556.1±41.5 µm (Spearman's rho=0.900, <0.001). No statistically significant difference in the new indices provided by the updated Corvis ST was detected among the three groups. Compared with control eyes, POAG eyes had a significantly reduced applanation time 2 after adjusting for OB-CCT and GAT IOP (=0.048).

Conclusion: Corvis b-IOP and CCT correlate well with GAT IOP and OB-CCT. b-IOP is not affected by CCT, which might be an advantage, especially in thick or thin corneas. Corvis ST yields shorter applanation time 2 measurements in patients with POAG, which might reflect altered corneal viscoelasticity.

Citing Articles

Corneal Biomechanical Measures for Glaucoma: A Clinical Approach.

Elhusseiny A, Scarcelli G, Saeedi O Bioengineering (Basel). 2023; 10(10).

PMID: 37892838 PMC: 10604716. DOI: 10.3390/bioengineering10101108.


Agreement of intraocular pressure measurement with Corvis ST, non-contact tonometer, and Goldmann applanation tonometer in children with ocular hypertension and related factors.

Li H, Chen Y, Lin F, Li S, Liu Q, Yin C Int J Ophthalmol. 2023; 16(10):1601-1607.

PMID: 37854370 PMC: 10559031. DOI: 10.18240/ijo.2023.10.07.


Progress of corneal morphological examination combined with biomechanical examination in preoperative screening for keratorefractive surgery.

Gui Y, Wang S, He Y, Zhang S, Zhang Y Indian J Ophthalmol. 2023; 71(6):2369-2378.

PMID: 37322646 PMC: 10417961. DOI: 10.4103/ijo.IJO_1377_22.

References
1.
Lee R, Chang R, Wong I, Lai J, Lee J, Singh K . Novel Parameter of Corneal Biomechanics That Differentiate Normals From Glaucoma. J Glaucoma. 2015; 25(6):e603-9. DOI: 10.1097/IJG.0000000000000284. View

2.
Vinciguerra R, Rehman S, Vallabh N, Batterbury M, Czanner G, Choudhary A . Corneal biomechanics and biomechanically corrected intraocular pressure in primary open-angle glaucoma, ocular hypertension and controls. Br J Ophthalmol. 2019; 104(1):121-126. PMC: 6922012. DOI: 10.1136/bjophthalmol-2018-313493. View

3.
Tai L, Khaw K, Ng C, Subrayan V . Central corneal thickness measurements with different imaging devices and ultrasound pachymetry. Cornea. 2012; 32(6):766-71. DOI: 10.1097/ICO.0b013e318269938d. View

4.
Sanchez-Barahona C, Bolivar G, Katsanos A, Teus M . Latanoprost treatment differentially affects intraocular pressure readings obtained with three different tonometers. Acta Ophthalmol. 2019; 97(8):e1112-e1115. DOI: 10.1111/aos.14170. View

5.
Aoki S, Kiuchi Y, Tokumo K, Fujino Y, Matsuura M, Murata H . Association between optic nerve head morphology in open-angle glaucoma and corneal biomechanical parameters measured with Corvis ST. Graefes Arch Clin Exp Ophthalmol. 2019; 258(3):629-637. DOI: 10.1007/s00417-019-04572-z. View