» Articles » PMID: 3521510

Recurrent Versus Residual Carotid Stenosis. Incidence Detected by Doppler Ultrasound

Overview
Journal Ann Surg
Specialty General Surgery
Date 1986 Jun 1
PMID 3521510
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

A sterile Doppler probe was used for intraoperative monitoring of the integrity of carotid endarterectomy of 125 vessels of 107 patients. Our objective was to reduce the contribution of residual carotid lesions to recurrent stenosis, which was evaluated by Doppler spectrum analysis in the early (less than 3 months, 66 arteries) and late (3-77 months, mean 32 months, 47 arteries) postoperative period. Intraoperative Doppler monitoring detected residual occlusion in six (4.8%) external carotid arteries and stenosis in ten (8.0%) internal carotid arteries permitting selected arteriography and correction of all significant lesions. Follow-up revealed one (1.5%) asymptomatic common carotid occlusion at 6 weeks and three (6.4%) asymptomatic internal carotid stenoses at 6, 10 and 25 months after operation. When compared to previous published experience, these results suggest that intraoperative correction of residual carotid obstruction detected by Doppler ultrasound may reduce the incidence of postoperative "recurrent" carotid stenosis.

Citing Articles

Sites of recurrence and long-term results of redo surgery.

Ballinger B, Money S, Chatman D, Bowen J, Ochsner J Ann Surg. 1997; 225(5):512-5; discussion 516-7.

PMID: 9193179 PMC: 1190787. DOI: 10.1097/00000658-199705000-00008.


A prospective study of vein patch angioplasty during carotid endarterectomy. Three-year results for 801 patients and 917 operations.

Hertzer N, Beven E, OHara P, Krajewski L Ann Surg. 1987; 206(5):628-35.

PMID: 3675024 PMC: 1493295. DOI: 10.1097/00000658-198711000-00013.


Does carotid restenosis predict an increased risk of late symptoms, stroke, or death?.

Bernstein E, Torem S, DILLEY R Ann Surg. 1990; 212(5):629-36.

PMID: 2241319 PMC: 1358192. DOI: 10.1097/00000658-199011000-00011.

References
1.
Rosental J, GASPAR M, MOVIUS H . Intraoperative arteriography in carotid thromboendarterectomy. Arch Surg. 1973; 106(6):806-8. DOI: 10.1001/archsurg.1973.01350180044014. View

2.
Seifert K, Blackshear Jr W . Continuous-wave Doppler in the intraoperative assessment of carotid endarterectomy. J Vasc Surg. 1985; 2(6):817-20. DOI: 10.1067/mva.1985.avs0020817. View

3.
Stoney R, String S . Recurrent carotid stenosis. Surgery. 1976; 80(6):705-10. View

4.
Towne J, Bernhard V . Vascular endoscopy --- an adjunct to carotid surgery. Stroke. 1977; 8(5):569-71. DOI: 10.1161/01.str.8.5.569. View

5.
French B, Rewcastle N . Recurrent stenosis at site of carotid endarterectomy. Stroke. 1977; 8(5):597-605. DOI: 10.1161/01.str.8.5.597. View