Abstract

Purpose: Hydrogen-methane breath tests are used to diagnose carbohydrate malabsorption and small intestinal bacterial overgrowth. The COVID-19 pandemic has driven the modification of procedures as breath tests are potentially aerosol-generating procedures. We assessed the effect of delayed analysis of breath samples, facilitating the at-home performance of breath testing.

Methods: Children provided two breath samples at every step of the lactose breath test. The samples were brought back to the clinic, and one set of samples was analyzed immediately. The second set was stored at room temperature and analyzed 1-4 days later.

Results: Out of the 73 "double" lactose breath tests performed at home, 33 (45.8%) were positive. The second samples were analyzed 20 to 117 hours after the first samples (41.7±24.3 hours). There was no significant difference in the hydrogen concentration between the first and second sets (Z=0.49, =0.62). This was not the case for methane, which had a significantly higher concentration in the second breath samples (Z=7.6).

Conclusion: Expired hydrogen levels remain stable in plastic syringes if preserved at room temperature for several days. On the other hand, the delayed analysis of methane appeared to be less reliable. Further research is needed to examine the impact of delayed analysis on methane and hydrogen concentrations.

Citing Articles

The impact of lactose intolerance diagnosis: costs, timing, and quality-of-life.

Dominici S, Donati N, Menabue S, Di Stefano M, Facioni M Intern Emerg Med. 2024; .

PMID: 39495353 DOI: 10.1007/s11739-024-03787-1.

References
1.
Rosado J, Solomons N . Storage of hydrogen breath-test samples in plastic syringes. Clin Chem. 1983; 29(3):583-4. View

2.
Solomons N, Viteri F, HAMILTON L . Application of a simple gas chromatographic technique for measuring breath hydrogen. J Lab Clin Med. 1977; 90(5):856-62. View

3.
Perman J, Barr R, WATKINS J . Sucrose malabsorption in children: noninvasive diagnosis by interval breath hydrogen determination. J Pediatr. 1978; 93(1):17-22. DOI: 10.1016/s0022-3476(78)80592-7. View

4.
Ghoshal U, Srivastava D, Misra A . A randomized double-blind placebo-controlled trial showing rifaximin to improve constipation by reducing methane production and accelerating colon transit: A pilot study. Indian J Gastroenterol. 2018; 37(5):416-423. DOI: 10.1007/s12664-018-0901-6. View

5.
Klompas M, Baker M, Rhee C . What Is an Aerosol-Generating Procedure?. JAMA Surg. 2020; 156(2):113-114. DOI: 10.1001/jamasurg.2020.6643. View