» Articles » PMID: 34993104

Comparison of Contrast-enhanced T1-weighted Imaging Using DANTE-SPACE, PETRA, and MPRAGE: a Clinical Evaluation of Brain Tumors at 3 Tesla

Overview
Specialty Radiology
Date 2022 Jan 7
PMID 34993104
Citations 2
Authors
Affiliations
Soon will be listed here.
Abstract

Background: We aimed to compare the performance of three contrast-enhanced T1-weighted three-dimensional (3D) magnetic resonance (MR) sequences to detect brain tumors at 3 Tesla. The three sequences were: (I) delay alternating with nutation for tailored excitation sampling perfection with application-optimized contrasts using different flip angle evolution (DANTE-SPACE), (II) pointwise encoding time reduction with radial acquisition (PETRA), and (III) magnetization-prepared rapid acquisition with gradient echo (MPRAGE).

Methods: This study involved 77 consecutive patients, including 34 patients with known primary brain tumors and 43 patients suspected of intracranial metastases. All patients underwent each of the three sequences with comparable spatial resolution and acquisition time post-injection. Signal-to-noise ratios (SNRs) for gray matter (GM) and white matter (WM), contrast-to-noise ratios (CNRs) for lesion/GM, lesion/WM, and GM/WM were quantitatively compared. Two radiologists determined the total number of enhancing lesions by consensus. Intraclass correlation coefficients (ICCs) between the two radiologists for metastases presence, qualitative ratings for image quality, and acoustic noise level of each sequence were assessed.

Results: Among the three sequences, SNRs and CNRs between lesions and surrounding parenchyma were highest using DANTE-SPACE, but CNR was the lowest with DANTE-SPACE. SNRs for PETRA images were significantly higher than those for MPRAGE (P<0.001). CNRs between lesions and surrounding parenchyma were similar for PETRA and MPRAGE (P>0.05). Significantly more brain metastases were detected with DANTE-SPACE (n=94) compared with MPRAGE (n=71) and PETRA (n=72). The ICCs were 0.964 for MPRAGE, 0.975 for PETRA, and 0.973 for DANTE-SPACE. Qualitative scores for lesion imaging using DANTE-SPACE were significantly higher than those obtained with PETRA and MPRAGE (P=0.002 and P=0.004, respectively). The acoustic noise level for PETRA (64.45 dB) was significantly lower than that for MPRAGE (78.27 dB, P<0.01) and DANTE-SPACE (80.18 dB, P<0.01).

Conclusions: PETRA achieves comparable detection of brain tumors with MPRAGE and is preferred for depicting osseous metastases and meningeal enhancement. DANTE-SPACE with blood vessel suppression showed improved detection of cerebral metastases compared with MPRAGE and PETRA, which could be helpful for the differential diagnosis of tumors.

Citing Articles

Diagnostics and Screening in Breast Cancer with Brain and Leptomeningeal Metastasis: A Review of the Literature.

Cohen-Nowak A, Hill V, Kumthekar P Cancers (Basel). 2024; 16(21).

PMID: 39518124 PMC: 11545036. DOI: 10.3390/cancers16213686.


Radiomics-Based Differentiation of Glioblastoma and Metastatic Disease: Impact of Different T1-Contrast-Enhanced Sequences on Radiomics Features and Model Performance.

Bathla G, Zamboni C, Larson N, Liu Y, Zhang H, Lee N AJNR Am J Neuroradiol. 2024; 46(2):321-329.

PMID: 39179298 PMC: 11878985. DOI: 10.3174/ajnr.A8470.


Optimization of workflow for detection of brain metastases at 3T: is a black-blood MTC prepared 3D T1 used alone robust enough to replace the combination of conventional 3D T1 and the black-blood 3D T1 MTC?.

Chkili S, Lefebvre Y, Chao S, Antonietta Bali M, Lemort M, Coquelet N Neuroradiology. 2023; 65(7):1133-1141.

PMID: 36995375 PMC: 10272253. DOI: 10.1007/s00234-023-03143-8.

References
1.
Kim D, Heo Y, Jeong H, Baek J, Han J, Lee J . Usefulness of the Delay Alternating with Nutation for Tailored Excitation Pulse with T1-Weighted Sampling Perfection with Application-Optimized Contrasts Using Different Flip Angle Evolution in the Detection of Cerebral Metastases: Comparison with.... AJNR Am J Neuroradiol. 2019; 40(9):1469-1475. PMC: 7048452. DOI: 10.3174/ajnr.A6158. View

2.
Xie Y, Yang Q, Xie G, Pang J, Fan Z, Li D . Improved black-blood imaging using DANTE-SPACE for simultaneous carotid and intracranial vessel wall evaluation. Magn Reson Med. 2015; 75(6):2286-94. PMC: 4706507. DOI: 10.1002/mrm.25785. View

3.
Wang J, Yarnykh V, Hatsukami T, Chu B, Balu N, Yuan C . Improved suppression of plaque-mimicking artifacts in black-blood carotid atherosclerosis imaging using a multislice motion-sensitized driven-equilibrium (MSDE) turbo spin-echo (TSE) sequence. Magn Reson Med. 2007; 58(5):973-81. DOI: 10.1002/mrm.21385. View

4.
Grodzki D, Jakob P, Heismann B . Ultrashort echo time imaging using pointwise encoding time reduction with radial acquisition (PETRA). Magn Reson Med. 2011; 67(2):510-8. DOI: 10.1002/mrm.23017. View

5.
Park J, Kim J, Yoo E, Lee H, Chang J, Kim E . Detection of small metastatic brain tumors: comparison of 3D contrast-enhanced whole-brain black-blood imaging and MP-RAGE imaging. Invest Radiol. 2011; 47(2):136-41. DOI: 10.1097/RLI.0b013e3182319704. View