Ultrasound Lesion Detectability As a Distance Between Probability Measures

Overview
Authors
Affiliations
Soon will be listed here.
Abstract

Lesion detectability (LD) quantifies how easily a lesion or target can be distinguished from the background. LD is commonly used to assess the performance of new ultrasound imaging methods. The contrast-to-noise ratio (CNR) is the most popular measure of LD; however, recent work has exposed its vulnerability to manipulations of dynamic range. The generalized CNR (gCNR) has been proposed as a robust histogram-based alternative that is invariant to such manipulations. Here, we identify key shortcomings of CNR and strengths of gCNR as LD metrics for modern beamformers. Using the measure theory, we pose LD as a distance between empirical probability measures (i.e., histograms) and prove that: 1) gCNR is equal to the total variation distance between probability measures and 2) gCNR is one minus the error rate of the ideal observer. We then explore several consequences of measure-theoretic LD in simulation studies. We find that histogram distances depend on bin selection that LD must be considered in the context of spatial resolution and that many histogram distances are invariant under measure-preserving isomorphisms of the sample space (e.g., dynamic range transformations). Finally, we provide a mathematical interpretation for why quantitative values such as contrast ratio (CR), CNR, and signal-to-noise ratio should not be compared between images with different dynamic ranges or underlying units and demonstrate how histogram matching can be used to reenable such quantitative comparisons.

Citing Articles

High-Resolution Power Doppler Using Null Subtraction Imaging.

Kou Z, Lowerison M, You Q, Wang Y, Song P, Oelze M IEEE Trans Med Imaging. 2024; 43(9):3060-3071.

PMID: 38557625 PMC: 11439488. DOI: 10.1109/TMI.2024.3383768.


Spatial Coherence Approaches to Distinguish Suspicious Mass Contents in Fundamental and Harmonic Breast Ultrasound Images.

Sharma A, Oluyemi E, Myers K, Ambinder E, Bell M IEEE Trans Ultrason Ferroelectr Freq Control. 2023; 71(1):70-84.

PMID: 37956000 PMC: 10851341. DOI: 10.1109/TUFFC.2023.3332207.


Methods for Enhancing the Robustness of the Generalized Contrast-to-Noise Ratio.

Schlunk S, Byram B IEEE Trans Ultrason Ferroelectr Freq Control. 2023; 70(8):831-842.

PMID: 37363856 PMC: 10481948. DOI: 10.1109/TUFFC.2023.3289157.

References
1.
Nguyen N, Prager R, Insana M . Improvements to ultrasonic beamformer design and implementation derived from the task-based analytical framework. J Acoust Soc Am. 2017; 141(6):4427. DOI: 10.1121/1.4985187. View

2.
Li P, Li M . Adaptive imaging using the generalized coherence factor. IEEE Trans Ultrason Ferroelectr Freq Control. 2003; 50(2):128-41. DOI: 10.1109/tuffc.2003.1182117. View

3.
Rodriguez-Molares A, Rindal O, Dhooge J, Masoy S, Austeng A, Bell M . The Generalized Contrast-to-Noise Ratio: A Formal Definition for Lesion Detectability. IEEE Trans Ultrason Ferroelectr Freq Control. 2019; 67(4):745-759. PMC: 8354776. DOI: 10.1109/TUFFC.2019.2956855. View

4.
Hyun D, Crowley A, Dahl J . Efficient Strategies for Estimating the Spatial Coherence of Backscatter. IEEE Trans Ultrason Ferroelectr Freq Control. 2016; 64(3):500-513. PMC: 5453518. DOI: 10.1109/TUFFC.2016.2634004. View

5.
Nguyen N, Abbey C, Insana M . Objective assessment of sonographic: quality II acquisition information spectrum. IEEE Trans Med Imaging. 2012; 32(4):691-8. DOI: 10.1109/TMI.2012.2231963. View