A method of reconstruction of single-photon emission computed tomographic (SPECT) images--distance-weighted filtered backprojection--is described. The quality of SPECT images using the standard filtered backprojection algorithm for data obtained from 360 degrees acquisitions is limited by the loss of spatial resolution with distance, and the attenuation and scatter inherent in the planar images used for reconstruction. In contrast to the standard techniques in which a reconstructed pixel receives equal contributions from 180 degrees opposed views, distance--weighted filtered backprojection applies a variable weighting factor to the data such that the pixel receives greater weight from the closest planar views. Data derived from hot and cold spot phantoms as well as from various clinical studies show better spatial and contrast resolution with this new technique compared with the conventional 360 degrees algorithm.

Citing Articles

Responsivity of the Striatal Dopamine System to Methylphenidate-A Within-Subject I-123-β-CIT-SPECT Study in Male Children and Adolescents With Attention-Deficit/Hyperactivity Disorder.

Aster H, Romanos M, Walitza S, Gerlach M, Muhlberger A, Rizzo A Front Psychiatry. 2022; 13:804730.

PMID: 35492708 PMC: 9046584. DOI: 10.3389/fpsyt.2022.804730.


High-resolution versus high-sensitivity SPECT imaging with geometric blurring compensation for various parallel-hole collimation geometries.

Zhang B, Zeng G IEEE Trans Inf Technol Biomed. 2010; 14(4):1121-7.

PMID: 20460211 PMC: 5292234. DOI: 10.1109/TITB.2010.2050145.


Effects of non-linear flow and spatial orientation on technetium-99m hexamethylpropylene amine oxime single-photon emission tomography.

Stapleton S, Caldwell C, Ehrlich L, Leonhardt C, Black S, Yaffe M Eur J Nucl Med. 1995; 22(9):1009-16.

PMID: 7588937 DOI: 10.1007/BF00808412.


Improved image quality with elliptical orbits and distance-weighted backprojection SPECT reconstruction.

Eisner R, Fajman W, Nowak D, Pettigrew R Ann Nucl Med. 1988; 2(2):107-10.

PMID: 3275107 DOI: 10.1007/BF03164954.


Three-dimensional display of myocardial perfusion: detection of ischemic lesions using a new image subtraction method.

Ferretti P, Borasi G, Salvo D, Serafini D, Versari A Eur J Nucl Med. 1990; 17(1-2):55-60.

PMID: 2083544 DOI: 10.1007/BF00819405.