First-In-Human Study in Cancer Patients Establishing the Feasibility of Oxygen Measurements in Tumors Using Electron Paramagnetic Resonance With the OxyChip
Overview
Authors
Affiliations
Objective: The overall objective of this clinical study was to validate an implantable oxygen sensor, called the 'OxyChip', as a clinically feasible technology that would allow individualized tumor-oxygen assessments in cancer patients prior to and during hypoxia-modification interventions such as hyperoxygen breathing.
Methods: Patients with any solid tumor at ≤3-cm depth from the skin-surface scheduled to undergo surgical resection (with or without neoadjuvant therapy) were considered eligible for the study. The OxyChip was implanted in the tumor and subsequently removed during standard-of-care surgery. Partial pressure of oxygen (pO) at the implant location was assessed using electron paramagnetic resonance (EPR) oximetry.
Results: Twenty-three cancer patients underwent OxyChip implantation in their tumors. Six patients received neoadjuvant therapy while the OxyChip was implanted. Median implant duration was 30 days (range 4-128 days). Forty-five successful oxygen measurements were made in 15 patients. Baseline pO values were variable with overall median 15.7 mmHg (range 0.6-73.1 mmHg); 33% of the values were below 10 mmHg. After hyperoxygenation, the overall median pO was 31.8 mmHg (range 1.5-144.6 mmHg). In 83% of the measurements, there was a statistically significant (p ≤ 0.05) response to hyperoxygenation.
Conclusions: Measurement of baseline pO and response to hyperoxygenation using EPR oximetry with the OxyChip is clinically feasible in a variety of tumor types. Tumor oxygen at baseline differed significantly among patients. Although most tumors responded to a hyperoxygenation intervention, some were non-responders. These data demonstrated the need for individualized assessment of tumor oxygenation in the context of planned hyperoxygenation interventions to optimize clinical outcomes.
Shaw M, Poncelet M, Banerjee D, Sierros K, Driesschaert B J Phys Chem B. 2025; 129(6):1846-1854.
PMID: 39883096 PMC: 11825265. DOI: 10.1021/acs.jpcb.4c06480.
Canavesi I, Viswakarma N, Khurana R, Epel B, Kuppusamy P, Pagel M Magn Reson Med. 2025; 93(5):2239-2249.
PMID: 39831522 PMC: 11893242. DOI: 10.1002/mrm.30418.
Surface dielectric resonator for in vivo EPR measurements.
Petryakov S, Kmiec M, Ubert C, Kassey V, Schaner P, Kuppusamy P J Magn Reson. 2024; 362:107690.
PMID: 38692250 PMC: 11102834. DOI: 10.1016/j.jmr.2024.107690.
Highly Sensitive Detection of Melanin in Melanomas Using Multi-harmonic Low Frequency EPR.
Wehbi M, Mignion L, Joudiou N, Harkemanne E, Gallez B Mol Imaging Biol. 2024; 26(3):484-494.
PMID: 38519805 PMC: 11211186. DOI: 10.1007/s11307-024-01911-3.
Hand-held electron spin resonance scanner for subcutaneous oximetry using OxyChip.
Almog N, Zgadzai O, Kuppusamy P, Zur Y, Baruch L, Machluf M Magn Reson Med. 2024; 92(1):430-439.
PMID: 38411265 PMC: 11055674. DOI: 10.1002/mrm.30066.