» Articles » PMID: 34516616

Current Clinical Practice in Patients with Cardiac Implantable Electronic Devices Undergoing Radiotherapy: a Literature Review

Overview
Journal Europace
Date 2021 Sep 13
PMID 34516616
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

Patients with cardiac implantable electronic devices (CIED) undergoing radiotherapy (RT) are more common due to the ageing of the population. With newer CIEDs' implementing the complementary metal-oxide semiconductor (CMOS) technology which allows the miniaturization of CIED, it is also more susceptible to RT. Effects of RT on CIED ranges from device interference, device operational/memory errors of permanent damage. These malfunctions can cause life-threatening clinical effects. Cumulative dose is not the only component of RT that causes CIED malfunction, as neutron use and dose rate effect also affects CIEDs. The management of this patient cohort in clinical practice is inconsistent due to the lack of a consistent guideline from manufacturers and physician specialty societies. Our review will focus on the current clinical practice and the recently updated guidelines of managing patients with CIED undergoing RT. We aim to simplify the evidence and provide a simple and easy to use guide based on the recent guidelines.

Citing Articles

Cardiac arrhythmias during and after thoracic irradiation for malignancies.

Heckmann M, Munster J, Finke D, Hund H, Schunn F, Debus J Cardiooncology. 2024; 10(1):81.

PMID: 39543778 PMC: 11562486. DOI: 10.1186/s40959-024-00277-3.


Radiation therapy for ventricular arrhythmias.

Liulu X, Balaji P, Barber J, De Silva K, Murray T, Hickey A J Med Imaging Radiat Oncol. 2024; 68(8):893-913.

PMID: 38698577 PMC: 11686466. DOI: 10.1111/1754-9485.13662.


Effects of Ionizing Radiation on Cardiac Implantable Electronic Devices (CIEDs) in Patients with Esophageal Cancer Undergoing Radiotherapy: A Pilot Study.

Uzun D, Salatzki J, Xynogalos P, Frey N, Debus J, Lang K Cancers (Basel). 2024; 16(3).

PMID: 38339306 PMC: 10854512. DOI: 10.3390/cancers16030555.


Impact of Radiotherapy on Malfunctions and Battery Life of Cardiac Implantable Electronic Devices in Cancer Patients.

Lisowski D, Lutyj P, Abazari A, Weick S, Traub J, Polat B Cancers (Basel). 2023; 15(19).

PMID: 37835524 PMC: 10571836. DOI: 10.3390/cancers15194830.


Increase in right ventricular lead pacing threshold following stereotactic ablative therapy for ventricular tachycardia.

Wijesuriya N, Galante J, Sisodia C, Whitaker J, Ahmad S, Rinaldi C HeartRhythm Case Rep. 2023; 9(8):555-559.

PMID: 37614389 PMC: 10444549. DOI: 10.1016/j.hrcr.2023.05.011.