» Articles » PMID: 34409572

One-tissue Compartment Model for Myocardial Perfusion Quantification with N-13 Ammonia PET Provides Matching Results: A Cross-comparison Between Carimas, FlowQuant, and PMOD

Abstract

Purpose: To cross-compare three software packages (SPs)-Carimas, FlowQuant, and PMOD-to quantify myocardial perfusion at global, regional, and segmental levels.

Materials And Methods: Stress N-13 ammonia PET scans of 48 patients with HCM were analyzed in three centers using Carimas, FlowQuant, and PMOD. Values agreed if they had an ICC > 0.75 and a difference < 20% of the median across all observers.

Results: When using 1TCM on the global level, the agreement was good, and the maximum difference between 1TCM MBF values was 17.2% (ICC = 0.83). On the regional level, the agreement was acceptable except in the LCx region (25.5% difference, ICC = 0.74) between FlowQuant and PMOD. Carimas-1TCM agreed well with PMOD-1TCM and FlowQuant-1TCM. Values obtained with FlowQuant-1TCM had a somewhat lesser agreement with PMOD-1TCM, especially at the segmental level.

Conclusions: The global and regional MBF values (with one exception) agree well between the different software packages. There is significant variability in segmental values, mainly located in the LCx region and segments. Out of the studied tools, Carimas can be used interchangeably with both PMOD and FlowQuant for 1TCM implementation on all levels-global, regional, and segmental.

Citing Articles

Inter-Software Reproducibility of Quantitative Values of Myocardial Blood Flow and Coronary Flow Reserve Acquired by [N]NH MPI PET/CT and the Effect of Motion Correction Tools.

Mendoza-Ibanez O, Slart R, Alexanderson-Rosas E, Martinez-Lucio T, van der Zant F, Knol R Diagnostics (Basel). 2025; 15(5).

PMID: 40075861 PMC: 11898590. DOI: 10.3390/diagnostics15050613.


The assessment of left ventricular volume and function in gated small animal F-FDG PET/CT imaging: a comparative study of three commercially available software tools.

Zacherl M, Simenhandra A, Lindner M, Bartenstein P, Todica A, Boening G EJNMMI Res. 2023; 13(1):75.

PMID: 37572238 PMC: 10423195. DOI: 10.1186/s13550-023-01026-w.


Measuring myocardial blood flow using dynamic myocardial perfusion SPECT: artifacts and pitfalls.

Mallet F, Poitrasson-Riviere A, Mariano-Goulart D, Agostini D, Manrique A J Nucl Cardiol. 2023; 30(5):2006-2017.

PMID: 36598748 DOI: 10.1007/s12350-022-03165-4.


Myocardial perfusion quantification with Rb-82 PET: good interobserver agreement of Carimas software on global, regional, and segmental levels.

Nesterov S, Deshayes E, Juarez-Orozco L, deKemp R, Sciagra R, Malaspina S Ann Nucl Med. 2022; 36(6):507-514.

PMID: 35192160 PMC: 9132838. DOI: 10.1007/s12149-022-01729-5.


How shall we judge a PET flow model?.

Johnson N, Gould K J Nucl Cardiol. 2021; 29(5):2551-2554.

PMID: 34561847 DOI: 10.1007/s12350-021-02805-5.

References
1.
Bland J, Altman D . Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986; 1(8476):307-10. View

2.
Cecchi F, Olivotto I, Gistri R, Lorenzoni R, Chiriatti G, Camici P . Coronary microvascular dysfunction and prognosis in hypertrophic cardiomyopathy. N Engl J Med. 2003; 349(11):1027-35. DOI: 10.1056/NEJMoa025050. View

3.
Camici P, Chiriatti G, Lorenzoni R, Bellina R, Gistri R, Italiani G . Coronary vasodilation is impaired in both hypertrophied and nonhypertrophied myocardium of patients with hypertrophic cardiomyopathy: a study with nitrogen-13 ammonia and positron emission tomography. J Am Coll Cardiol. 1991; 17(4):879-86. DOI: 10.1016/0735-1097(91)90869-b. View

4.
DeGrado T, Hanson M, Turkington T, Delong D, Brezinski D, Vallee J . Estimation of myocardial blood flow for longitudinal studies with 13N-labeled ammonia and positron emission tomography. J Nucl Cardiol. 1996; 3(6 Pt 1):494-507. DOI: 10.1016/s1071-3581(96)90059-8. View

5.
Nesterov S, Deshayes E, Sciagra R, Settimo L, Declerck J, Pan X . Quantification of myocardial blood flow in absolute terms using (82)Rb PET imaging: the RUBY-10 Study. JACC Cardiovasc Imaging. 2014; 7(11):1119-1127. PMC: 4260449. DOI: 10.1016/j.jcmg.2014.08.003. View