Whole Body Planar and 3D Quantitative Imaging After Lu-DOTATATE on CZT SPECT/CT Device with MEHRS Collimator. First Report
Overview
Authors
Affiliations
Lutetium-177 (Lu)-based post-therapeutic imaging allows visualization of treated lesions andabsorbed dose measurement. There is an increasing number of cadmium-zinc-telluride (CZT) gamma-cameras in nuclear medicine departments but until now these devices were not adapted to the medium-energy emission of Lu photons. We present here in the first reported images acquired with a new collimator designed for CZT gamma-camera compared to a conventional sodium iodide (NaI) (Tl) gamma-camera. Post-therapeutic Lu-DOTATATE imaging on a CZT device with a medium energy high resolution (MEHRS)-collimator are promising and support the widespread of both Lu-based peptide-receptor radionuclide therapy (PRRT) and CZT gamma-cameras.
Sweet C, Shmuel N, Shoaf J, Stoecklein M, Muthukrishnan A, Stern E Nucl Med Mol Imaging. 2024; 58(1):1-8.
PMID: 38250182 PMC: 10796310. DOI: 10.1007/s13139-023-00825-2.