Dosimetry on First Clinical Dark-field Chest Radiography

Abstract

Purpose: The purpose of this study was to evaluate the dose characteristic for patient examinations at the first clinical X-ray dark-field chest radiography system and to determine whether the effective patient dose is within a clinically acceptable dose range.

Methods: A clinical setup for grating-based dark-field chest radiography was constructed and commissioned, operating at a tube voltage of 70 kVp. Thermoluminescent dosimeter (TLD) measurements were conducted using an anthropomorphic phantom modeling the reference person to obtain a conversion coefficient relating dose area product (DAP) to effective patient dose at the dark-field system. For 92 patients, the DAP values for posterior-anterior measurements were collected at the dark-field system. Using the previously determined conversion coefficient, the effective dose was calculated.

Results: A reference person, modeled by an anthropomorphic phantom, receives an effective dose of 35 µSv. For the examined patients, a mean effective dose of 39 µSv was found.

Conclusions: The effective dose at the clinical dark-field radiography system, generating both attenuation and dark-field images, is within the range of reported standard dose values for chest radiography.

Citing Articles

Comparison of dark-field chest radiography and CT for the assessment of COVID-19 pneumonia.

Gassert F, Bast H, Urban T, Frank M, Gassert F, Willer K Front Radiol. 2025; 4:1487895.

PMID: 39877725 PMC: 11772474. DOI: 10.3389/fradi.2024.1487895.


Simulated low-dose dark-field radiography for detection of COVID-19 pneumonia.

Schick R, Bast H, Frank M, Urban T, Koehler T, Gassert F PLoS One. 2024; 19(12):e0316104.

PMID: 39729472 PMC: 11676568. DOI: 10.1371/journal.pone.0316104.


A compact X-ray source via fast microparticle streams.

Behling R, Hulme C, Poludniowski G, Tolias P, Danielsson M Commun Eng. 2024; 3(1):171.

PMID: 39548184 PMC: 11568210. DOI: 10.1038/s44172-024-00323-z.


Dark-field radiography for the detection of bone microstructure changes in osteoporotic human lumbar spine specimens.

Rischewski J, Gassert F, Urban T, Hammel J, Kufner A, Braun C Eur Radiol Exp. 2024; 8(1):125.

PMID: 39495387 PMC: 11534944. DOI: 10.1186/s41747-024-00524-3.


[Dark-field imaging and computed tomography : Novel X-ray-based contrast imaging modality with great promise for pulmonary imaging].

Pfeiffer F, Willer K, Viermetz M, Pfeiffer D Radiologie (Heidelb). 2023; 63(7):513-522.

PMID: 37341743 DOI: 10.1007/s00117-023-01161-4.