Technical Note: Practical Implementation Strategies of Cycloidal Computed Tomography

Overview
Journal Med Phys
Specialty Biophysics
Date 2021 Jun 25
PMID 34169514
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: Cycloidal computed tomography is a novel imaging concept which combines a highly structured x-ray beam, offset lateral under-sampling, and mathematical data recovery to obtain high-resolution images efficiently and flexibly, even with relatively large source focal spots and detector pixels. The method reduces scanning time and, potentially, delivered dose compared to other sampling schemes. This study aims to present and discuss several implementation strategies for cycloidal computed tomography (CT) in order to increase its ease of use and facilitate uptake within the imaging community.

Methods: The different implementation strategies presented are step-and-shoot, continuous unidirectional, continuous back-and-forth, and continuous pixel-wise scanning. In step-and-shoot scans the sample remains stationary while frames are acquired, whereas in all other cases the sample moves through the scanner continuously. The difference between the continuous approaches is the trajectory by which the sample moves within the field of view.

Results: All four implementation strategies are compatible with a standard table-top x-ray setup. With the experimental setup applied here, step-and-shoot acquisitions yield the best spatial resolution (around 30 µm), but are the most time-consuming (1.4 h). Continuous unidirectional and back-and-forth images have resolution between 30 and 40 µm, and are faster (35 min). Continuous pixel-wise images are equally time-efficient, although technical challenges caused a small loss in image quality with a resolution of about 50 µm.

Conclusion: The authors show that cycloidal CT can be implemented in a variety of ways with high quality results. They believe this posits cycloidal CT as a powerful imaging alternative to more time-consuming and less flexible methods in the field.

Citing Articles

Phase contrast micro-CT with adjustable in-slice spatial resolution at constant magnification.

Zekavat A, Lioliou G, Roche I Morgo O, Maughan Jones C, Galea G, Maniou E Phys Med Biol. 2024; 69(10).

PMID: 38631365 PMC: 11075091. DOI: 10.1088/1361-6560/ad4000.


Edge-illumination spectral phase-contrast tomography.

Brombal L, Arfelli F, Brun F, Di Trapani V, Endrizzi M, Menk R Phys Med Biol. 2024; 69(7).

PMID: 38471186 PMC: 10991267. DOI: 10.1088/1361-6560/ad3328.


Reducing radiation dose for NN-based COVID-19 detection in helical chest CT using real-time monitored reconstruction.

Bulatov K, Ingacheva A, Gilmanov M, Chukalina M, Nikolaev D, Arlazarov V Expert Syst Appl. 2023; 229:120425.

PMID: 37215381 PMC: 10176897. DOI: 10.1016/j.eswa.2023.120425.


PEPI Lab: a flexible compact multi-modal setup for X-ray phase-contrast and spectral imaging.

Brombal L, Arfelli F, Menk R, Rigon L, Brun F Sci Rep. 2023; 13(1):4206.

PMID: 36918574 PMC: 10014955. DOI: 10.1038/s41598-023-30316-5.


Utility of knife-edge position tracking in cycloidal computed tomography.

Roche I Morgo O, Aleksejev J, Astolfo A, Savvidis S, Gerli M, Cipiccia S Opt Express. 2022; 30(24):43209-43222.

PMID: 36523024 PMC: 9765405. DOI: 10.1364/OE.470798.


References
1.
Momose A . Development toward high-resolution X-ray phase imaging. Microscopy (Oxf). 2017; 66(3):155-166. DOI: 10.1093/jmicro/dfx013. View

2.
Wang J, Fleischmann D . Improving Spatial Resolution at CT: Development, Benefits, and Pitfalls. Radiology. 2018; 289(1):261-262. DOI: 10.1148/radiol.2018181156. View

3.
Grimes J, Duan X, Yu L, Halaweish A, Haag N, Leng S . The influence of focal spot blooming on high-contrast spatial resolution in CT imaging. Med Phys. 2015; 42(10):6011-20. DOI: 10.1118/1.4931053. View

4.
Munro P, Ignatyev K, Speller R, Olivo A . Phase and absorption retrieval using incoherent X-ray sources. Proc Natl Acad Sci U S A. 2012; 109(35):13922-7. PMC: 3435200. DOI: 10.1073/pnas.1205396109. View

5.
Diemoz P, Endrizzi M, Bravin A, Robinson I, Olivo A . Sensitivity of edge illumination X-ray phase-contrast imaging. Philos Trans A Math Phys Eng Sci. 2014; 372(2010):20130128. PMC: 3900038. DOI: 10.1098/rsta.2013.0128. View