Impact of a Formula Combining Local Impedance and Conventional Parameters on Lesion Size Prediction

Abstract

Background: Although ablation energy (AE) and force-time integral (FTI) are well-known active predictors of lesion characteristics, these parameters do not reflect passive tissue reactions during ablation, which may instead be represented by drops in local impedance (LI). This study aimed to investigate if additional LI data improves predicting lesion characteristics and steam pops.

Methods: RF applications at a range of powers (30 W, 40 W, and 50 W), contact forces (8 g, 15 g, 25 g, and 35 g), and durations (10-180 s) using perpendicular/parallel catheter orientations were performed in excised porcine hearts (N = 30). The correlation between AE, FTI, and lesion characteristics was examined, and the impact of LI (%LI drop (%LID) defined by the ΔLI divided by the initial LI) was additionally assessed.

Results: Three hundred seventy-five lesions without steam pops were examined. Ablation energy (W × s) and FTI (g × s) showed a positive correlation with lesion depth (ρ = 0.824:P < 0.0001 and ρ = 0.708:P < 0.0001), surface area (ρ = 0.507:P < 0.0001 and ρ = 0.562:P < 0.0001), and volume (ρ = 0.807:P < 0.0001 and ρ = 0.685:P < 0.0001). %LID also showed a positive correlation individually with lesion depth (ρ = 0.643:P < 0.0001), surface area (ρ = 0.547:P < 0.0001), and volume (ρ = 0.733:P < 0.0001). However, the combined indices of AE × %LID (AE multiplied by %LID) and FTI × %LID (FTI multiplied by %LID) provided significantly stronger correlation with lesion depth (ρ = 0.834:P < 0.0001 and ρ = 0.809:P < 0.0001), surface area (ρ = 0.529:P < 0.0001 and ρ = 0.656:P < 0.0001), and volume (ρ = 0.864:P < 0.0001 and ρ = 0.838:P < 0.0001). This tendency was observed regardless of the catheter placement (parallel/perpendicular). AE (P = 0.02) and %LID (P = 0.002) independently remained as significant predictors to predict steam pops (N = 27). However, the AE × %LID did not increase the predictive power of steam pops compared to the AE alone.

Conclusion: LI, when combined with conventional parameters (AE and FTI), may provide stronger correlation with lesion characteristics.

Citing Articles

An optimized approach for increasing lesion size in temperature-controled setting using a catheter with a surface thermocouple and efficient irrigation.

Takigawa M, Yamaguchi J, Goya M, Iwakawa H, Yamamoto T, Amemiya M J Arrhythm. 2024; 40(3):536-551.

PMID: 38939786 PMC: 11199823. DOI: 10.1002/joa3.13040.


Impact of contact force on the lesion characteristics of very high-power short-duration ablation using a QDOT-MICRO catheter.

Yamaguchi J, Takigawa M, Goya M, Martin C, Negishi M, Yamamoto T J Arrhythm. 2024; 40(2):247-255.

PMID: 38586837 PMC: 10995585. DOI: 10.1002/joa3.12992.


Impact of tip design and thermocouple location on the efficacy and safety of radiofrequency application.

Yamaguchi J, Takigawa M, Goya M, Martin C, Amemiya M, Yamamoto T J Interv Card Electrophysiol. 2022; 66(4):885-896.

PMID: 35451737 DOI: 10.1007/s10840-022-01219-8.


Comparison of two catheters measuring local impedance: local impedance variation vs lesion characteristics and steam pops.

Amemiya M, Takigawa M, Goya M, Martin C, Anzai T, Takahashi K J Interv Card Electrophysiol. 2022; 65(2):419-428.

PMID: 35438394 DOI: 10.1007/s10840-022-01214-z.

References
1.
Thiagalingam A, dAvila A, Foley L, Guerrero J, Lambert H, Leo G . Importance of catheter contact force during irrigated radiofrequency ablation: evaluation in a porcine ex vivo model using a force-sensing catheter. J Cardiovasc Electrophysiol. 2010; 21(7):806-11. DOI: 10.1111/j.1540-8167.2009.01693.x. View