To Compare Central Corneal Thickness Measurements Obtained by Pentacam with Those Obtained by IOLMaster 700, Cirrus Anterior Segment Optical Coherence Tomography and Tomey Specular Microscopy in Normal Healthy Eyes

Overview
Specialty Ophthalmology
Date 2021 Jun 19
PMID 34146013
Citations 4
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To compare central corneal thickness measurements obtained by Pentacam with those obtained by IOL Master 700, Cirrus Anterior segment optical coherence tomography and Tomey Specular microscopy in normal healthy eyes.

Methods: Two hundred and six eyes of healthy subjects were included in the study. Each subject was assessed by four different methods of measuring central corneal thickness using Pentacam, IOL Master 700, Cirrus AS-OCT and Tomey Specular microscopy by a single examiner.

Results: The mean CCT [± standard deviation (SD)] for Pentacam, IOL Master 700, Cirrus AS-OCT and Tomey Specular microscopy were Pentacam (Oculus), AS-OCT (Cirrus), IOL Master 700 and Specular microscopy (Tomey) were 523.75 (±27.75), 525.29 (±28.81),517.13 (±28.43) and 512.82 (±27.60) μm, respectively. All the means were significantly different from one another (P < 0.000). The differences between pairs of mean central corneal thickness (CCT) for Pentacam and IOL Master, Pentacam and anterior segment- optical coherence tomography (AS-OCT), and Pentacam and Specular microscopy are statistically significant. Bland-Altman plots showed that pentacam and IOL Master 700 have the closest agreement, followed by AS-OCT. Specular microscopy was found to have the poorest agreement with Pentacam.

Conclusion: We found that CCT measurements of Pentacam did not correlate with measurements of IOL Master, or AS-OCT or Specular microscopy. In clinical practice, the devices analyzed should not be used interchangeably due to low agreement regarding CCT values.

Citing Articles

Interocular Symmetry and Intermachine Reproducibility of Optic Disc and Macular Parameters Measured by Two Different Models of Optical Coherence Tomography.

Firdaus S, Pereira L, Yang G, Huang-Link Y Clin Ophthalmol. 2024; 18:2397-2406.

PMID: 39221077 PMC: 11363946. DOI: 10.2147/OPTH.S465360.


Comparison of ocular biometric parameters between two swept-source optical coherence tomography devices and Scheimpflug tomography in patients with cataract.

Ma S, Li C, Sun J, Yang J, Wen K, Chen X Int J Ophthalmol. 2024; 17(8):1437-1446.

PMID: 39156774 PMC: 11286440. DOI: 10.18240/ijo.2024.08.08.


Comparison of corneal measurements in normal and keratoconus eyes using Anterior Segment Optical Coherence Tomography (AS-OCT) and Pentacam HR topographer.

Said O, Kamal M, Tawfik S, Saif A BMC Ophthalmol. 2023; 23(1):194.

PMID: 37138239 PMC: 10155327. DOI: 10.1186/s12886-023-02946-w.


Monitoring of central corneal thickness after phacoemulsification-comparison of statical and rotating Scheimpflug pachymetry, and spectral-domain OCT.

Handzel D, Meyer C, Wegener A Int J Ophthalmol. 2022; 15(8):1266-1272.

PMID: 36017056 PMC: 9358173. DOI: 10.18240/ijo.2022.08.07.

References
1.
BarKana Y, Gerber Y, Elbaz U, Schwartz S, Ken-Dror G, Avni I . Central corneal thickness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound pachymetry. J Cataract Refract Surg. 2005; 31(9):1729-35. DOI: 10.1016/j.jcrs.2005.03.058. View

2.
Yap T, Archer T, Gobbe M, Reinstein D . Comparison of Central Corneal Thickness Between Fourier-Domain OCT, Very High-Frequency Digital Ultrasound, and Scheimpflug Imaging Systems. J Refract Surg. 2016; 32(2):110-6. DOI: 10.3928/1081597X-20151223-01. View

3.
Chen S, Huang J, Wen D, Chen W, Huang D, Wang Q . Measurement of central corneal thickness by high-resolution Scheimpflug imaging, Fourier-domain optical coherence tomography and ultrasound pachymetry. Acta Ophthalmol. 2010; 90(5):449-55. DOI: 10.1111/j.1755-3768.2010.01947.x. View

4.
Wheeler N, Morantes C, Kristensen R, Pettit T, Lee D . Reliability coefficients of three corneal pachymeters. Am J Ophthalmol. 1992; 113(6):645-51. DOI: 10.1016/s0002-9394(14)74788-9. View

5.
Fujioka M, Nakamura M, Tatsumi Y, Kusuhara A, Maeda H, Negi A . Comparison of Pentacam Scheimpflug camera with ultrasound pachymetry and noncontact specular microscopy in measuring central corneal thickness. Curr Eye Res. 2007; 32(2):89-94. DOI: 10.1080/02713680601115010. View