Development of Clinical Application Program for Radiotherapy Induced Cancer Risk Calculation Using Monte Carlo Engine in Volumetric-modulated Arc Therapy

Overview
Journal Radiat Oncol
Publisher Biomed Central
Specialties Oncology
Radiology
Date 2021 Jun 13
PMID 34118968
Citations 4
Authors
Affiliations
Soon will be listed here.
Abstract

Background: The purpose of this study is to develop a clinical application program that automatically calculates the effect for secondary cancer risk (SCR) of individual patient. The program was designed based on accurate dose calculations using patient computed tomography (CT) data and Monte Carlo engine. Automated patient-specific evaluation program was configured to calculate SCR.

Methods: The application program is designed to re-calculate the beam sequence of treatment plan using the Monte Carlo engine and patient CT data, so it is possible to accurately calculate and evaluate scatter and leakage radiation, difficult to calculate in TPS. The Monte Carlo dose calculation system was performed through stoichiometric calibration using patient CT data. The automatic SCR evaluation program in application program created with a MATLAB was set to analyze the results to calculate SCR. The SCR for organ of patient was calculated based on Biological Effects of Ionizing Radiation (BEIR) VII models. The program is designed to sequentially calculate organ equivalent dose (OED), excess absolute risk (EAR), excess relative risk (ERR), and the lifetime attributable risk (LAR) in consideration of 3D dose distribution analysis. In order to confirm the usefulness of the developed clinical application program, the result values from clinical application program were compared with the manual calculation method used in the previous study.

Results: The OED values calculated in program were calculated to be at most approximately 13.3% higher than results in TPS. The SCR result calculated by the developed clinical application program showed a maximum difference of 1.24% compared to the result of the conventional manual calculation method. And it was confirmed that EAR, ERR and LAR values can be easily calculated by changing the biological parameters.

Conclusions: We have developed a patient-specific SCR evaluation program that can be used conveniently in the clinic. The program consists of a Monte Carlo dose calculation system for accurate calculation of scatter and leakage radiation and a patient-specific automatic SCR evaluation program using 3D dose distribution. The clinical application program that improved the disadvantages of the existing process can be used as an index for evaluating a patient treatment plan.

Citing Articles

A framework for in-field and out-of-field patient specific secondary cancer risk estimates from treatment plans using the TOPAS Monte Carlo system.

Meyer I, Peters N, Tamborino G, Lee H, Bertolet A, Faddegon B Phys Med Biol. 2024; 69(16).

PMID: 39019051 PMC: 11345907. DOI: 10.1088/1361-6560/ad64b6.


A generalization performance study on the boosting radiotherapy dose calculation engine based on super-resolution.

Wang Y, Liu Y, Bai Y, Zhou Q, Xu S, Pang X Z Med Phys. 2023; 34(2):208-217.

PMID: 36631314 PMC: 11639714. DOI: 10.1016/j.zemedi.2022.10.006.


Experimental Validation of an Analytical Program and a Monte Carlo Simulation for the Computation of the Far Out-of-Field Dose in External Beam Photon Therapy Applied to Pediatric Patients.

De Saint-Hubert M, Suesselbeck F, Vasi F, Stuckmann F, Rodriguez M, Dabin J Front Oncol. 2022; 12:882506.

PMID: 35875147 PMC: 9300838. DOI: 10.3389/fonc.2022.882506.


Assessment of dose perturbations for metal stent in photon and proton radiotherapy plans for hepatocellular carcinoma.

Lee B, Cho S, Park H, Kang S, Kim J, Chung J Radiat Oncol. 2022; 17(1):125.

PMID: 35842636 PMC: 9288675. DOI: 10.1186/s13014-022-02100-8.

References
1.
Ruben J, Lancaster C, Jones P, Smith R . A comparison of out-of-field dose and its constituent components for intensity-modulated radiation therapy versus conformal radiation therapy: implications for carcinogenesis. Int J Radiat Oncol Biol Phys. 2010; 81(5):1458-64. DOI: 10.1016/j.ijrobp.2010.08.008. View

2.
Newhauser W, Durante M . Assessing the risk of second malignancies after modern radiotherapy. Nat Rev Cancer. 2011; 11(6):438-48. PMC: 4101897. DOI: 10.1038/nrc3069. View

3.
Kourinou K, Mazonakis M, Lyraraki E, Stratakis J, Damilakis J . Scattered dose to radiosensitive organs and associated risk for cancer development from head and neck radiotherapy in pediatric patients. Phys Med. 2012; 29(6):650-5. DOI: 10.1016/j.ejmp.2012.08.001. View

4.
Berris T, Mazonakis M, Stratakis J, Tzedakis A, Fasoulaki A, Damilakis J . Calculation of organ doses from breast cancer radiotherapy: a Monte Carlo study. J Appl Clin Med Phys. 2013; 14(1):4029. PMC: 5713920. DOI: 10.1120/jacmp.v14i1.4029. View

5.
Mesbahi A, Seyednejad F, Gasemi-Jangjoo A . Estimation of organs doses and radiation-induced secondary cancer risk from scattered photons for conventional radiation therapy of nasopharynx: a Monte Carlo study. Jpn J Radiol. 2010; 28(5):398-403. DOI: 10.1007/s11604-010-0432-2. View