Severe Corrosion of Modular Dual Mobility Acetabular Components Identified During Revision Total Hip Arthroplasty

Overview
Publisher Elsevier
Date 2021 Jun 7
PMID 34095398
Citations 6
Authors
Affiliations
Soon will be listed here.
Abstract

There has been a significant increase in the use of modular dual mobility (MDM) acetabular cups for primary and revision total hip arthroplasty (THA) secondary to decreased dislocation rate and increased impingement-free range of motion. Mating of dissimilar metals in THA can result in mechanically assisted crevice corrosion, with increased serum metal ion levels and potentially adverse local tissue reaction that can lead to revision surgery. In this case report, we present a patient who had THA using MDM components, and his follow-up laboratory testing showed asymptomatic elevated serum cobalt level. The patient subsequently developed an intraprosthetic dislocation that required revision surgery. At the time of revision, significant corrosion was noted on the backside of the modular acetabular liner and the inner surface of the titanium shell. Serum cobalt levels returned to normal after revision surgery. This case suggests that mechanically assisted crevice corrosion at the modular surface of MDM components does occur. In addition, we believe surgeons should judiciously use MDM technology only for patients at significantly increased risk of dislocation after THA.

Citing Articles

Corrosion of Modular Dual-Mobility Acetabular Components Leading to Acetabular Bone Loss and Protrusio.

Yang J, Dasari S, Chansky H, Hernandez N Arthroplast Today. 2024; 30:101543.

PMID: 39502436 PMC: 11536004. DOI: 10.1016/j.artd.2024.101543.


Two-Year Outcomes of Novel Dual-Mobility Implant in Primary Total Hip Arthroplasty.

Schaffler B, Raymond H, Black C, Habibi A, Ehlers M, Duncan S Adv Orthop. 2024; 2024:4125965.

PMID: 38264013 PMC: 10805547. DOI: 10.1155/2024/4125965.


No Change in Serum Metal Ions Levels After Primary Total Hip Replacement With an Additively Manufactured Dual Mobility Acetabular Construct.

Alpaugh K, Mishu M, Westrich G Arthroplast Today. 2022; 17:132-135.

PMID: 36091082 PMC: 9449544. DOI: 10.1016/j.artd.2022.07.019.


Dual-mobility versus Fixed-bearing in Primary Total Hip Arthroplasty: Outcome Comparison.

Singh V, Loloi J, Macaulay W, Hepinstall M, Schwarzkopf R, Aggarwal V Hip Pelvis. 2022; 34(2):96-105.

PMID: 35800126 PMC: 9204238. DOI: 10.5371/hp.2022.34.2.96.


Neurologic Dysfunction Associated With Mechanically Assisted Crevice Corrosion and Elevated Cobalt Ion Levels After Total Hip Arthroplasty.

Yan B, Bini S Arthroplast Today. 2021; 11:217-221.

PMID: 34646925 PMC: 8501461. DOI: 10.1016/j.artd.2021.09.002.


References
1.
Jacobs J, Cooper H, Urban R, Wixson R, Della Valle C . What do we know about taper corrosion in total hip arthroplasty?. J Arthroplasty. 2014; 29(4):668-9. DOI: 10.1016/j.arth.2014.02.014. View

2.
Batailler C, Fary C, Verdier R, Aslanian T, Caton J, Lustig S . The evolution of outcomes and indications for the dual-mobility cup: a systematic review. Int Orthop. 2016; 41(3):645-659. DOI: 10.1007/s00264-016-3377-y. View

3.
Cooper H, Della Valle C, Berger R, Tetreault M, Paprosky W, Sporer S . Corrosion at the head-neck taper as a cause for adverse local tissue reactions after total hip arthroplasty. J Bone Joint Surg Am. 2013; 94(18):1655-61. PMC: 3444948. DOI: 10.2106/jbjs.k.01352. View

4.
Lawrence H, Deehan D, Holland J, Kirby J, Tyson-Capper A . The immunobiology of cobalt: demonstration of a potential aetiology for inflammatory pseudotumours after metal-on-metal replacement of the hip. Bone Joint J. 2014; 96-B(9):1172-7. DOI: 10.1302/0301-620X.96B9.33476. View

5.
Agne M, Underwood R, Kocagoz S, MacDonald D, Day J, Parvizi J . Is there material loss at the backside taper in modular CoCr acetabular liners?. Clin Orthop Relat Res. 2014; 473(1):275-85. PMC: 4390963. DOI: 10.1007/s11999-014-3982-2. View