Background: The Silk Vista Baby (SVB) flow diverter (FDS) is the only FDS deliverable via a 0.017 inch microcatheter and is specifically designed for the distal vasculature. We sought to evaluate the safety and efficacy of the SVB.

Materials And Methods: We performed a retrospective review to identify SVB cases at 4 tertiary neurosurgical centres within the U.K. Clinical, procedural, angiographic and follow-up data were collected.

Results: We identified 60 patients (35 female, 58%) of average age 54 ± 10.5 (range 30-72) with 61 aneurysms, 50 (81.9%) located in the anterior circulation. The majority of the aneurysms treated were unruptured (46, 75.4%) and saccular (46, 75.4%). Dome size was 6.2 ± 6.2 mm (range 1-36mm) and parent vessel diameter was 2.3 ± 0.4 mm (range 1.2-3.3 mm).An average number of 1.07 devices were implanted. Coils or other devices were implanted in 14 aneurysms (23.3%). At last angiographic follow-up (n = 55), 7.5 ± 4.2 months post-procedure, 32 aneurysms (57.1%) were graded as RRC I, 7 (12.5%) RRC II, and 17 RRC III (30.4%).Clinical complications, excluding death, were seen in 4 patients (6.8%) including 1 delayed aneurysm rupture and 3 symptomatic ischaemic events. Only one patient had permanent morbidity (mRS 1). 3 patients died during follow-up (5.1%); 2 deaths were related to the aneurysms (3.4%) - one ruptured dissecting MCA aneurysm, and one giant partially thrombosed posterior circulation aneurysm. 93% of patients were mRS ≤ 2 at last follow-up.

Conclusion: The SVB has high rates of technical success and an acceptable safety profile. Distal aneurysms may occlude slower due to relative oversizing of the devices.

Citing Articles

Endovascular treatment of intracranial aneurysms with silk vista baby flow diverter: A systematic review and meta-analysis.

Hajikarimloo B, Mohammadzadeh I, Tos S, Hashemi R, Habibi M, Hasanzade A Neuroradiol J. 2025; :19714009251324324.

PMID: 40016880 PMC: 11869223. DOI: 10.1177/19714009251324324.


Technical success, procedural safety, and efficacy of the Silk Vista Baby in the treatment of cerebral aneurysms over a mid-to-long-term follow-up.

Rodriguez-Fernandez C, Ruiz-Garcia P, Garcia-Sanchez M, Manrique-Zegarra M, Toledano-Illan C, Escartin J Front Neurol. 2024; 15:1369443.

PMID: 38638309 PMC: 11025534. DOI: 10.3389/fneur.2024.1369443.


Endovascular treatment of challenging aneurysms with FRED Jr flow diverter stents: a single-center experience.

Sayin B, Senol Y, Daglioglu E, Ozbakir M, Orhan G, Akmangit I Jpn J Radiol. 2022; 41(3):322-334.

PMID: 36315360 PMC: 9619020. DOI: 10.1007/s11604-022-01354-2.


Patient outcomes after treatment of brain aneurysm in small diameter vessels with the silk vista baby flow diverter: A systematic review.

Hanel R, Cortez G, Benalia V, Sheffels E, Sutphin D, Pederson J Interv Neuroradiol. 2022; 30(1):5-13.

PMID: 35392703 PMC: 10956459. DOI: 10.1177/15910199221091645.

References
1.
Yeomans J, Sandu L, Sastry A . Pipeline Flex embolisation device with Shield Technology for the treatment of patients with intracranial aneurysms: periprocedural and 6 month outcomes. Neuroradiol J. 2020; 33(6):471-478. PMC: 7788678. DOI: 10.1177/1971400920966749. View

2.
Pistocchi S, Blanc R, Bartolini B, Piotin M . Flow diverters at and beyond the level of the circle of willis for the treatment of intracranial aneurysms. Stroke. 2012; 43(4):1032-8. DOI: 10.1161/STROKEAHA.111.636019. View

3.
Pierot L, Soize S, Cappucci M, Manceau P, Riva R, Eker O . Surface-modified flow diverter p48-MW-HPC: Preliminary clinical experience in 28 patients treated in two centers. J Neuroradiol. 2020; 48(3):195-199. DOI: 10.1016/j.neurad.2020.11.006. View

4.
Bhogal P, Wong K, Uff C, Wadley J, Makalanda H . The Silk Vista Baby: Initial experience and report of two cases. Interv Neuroradiol. 2019; 25(5):530-538. PMC: 6777101. DOI: 10.1177/1591019919839196. View

5.
Shapiro M, Raz E, Becske T, Nelson P . Variable porosity of the pipeline embolization device in straight and curved vessels: a guide for optimal deployment strategy. AJNR Am J Neuroradiol. 2013; 35(4):727-33. PMC: 7965797. DOI: 10.3174/ajnr.A3742. View