CT-less Electron Radiotherapy Simulation and Planning with a Consumer 3D Camera

Overview
Date 2021 May 27
PMID 34042253
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: Electron radiation therapy dose distributions are affected by irregular body surface contours. This study investigates the feasibility of three-dimensional (3D) cameras to substitute for the treatment planning computerized tomography (CT) scan by capturing the body surfaces to be treated for accurate electron beam dosimetry.

Methods: Dosimetry was compared for six electron beam treatments to the nose, toe, eye, and scalp using full CT scan, CT scan with Hounsfield Unit (HU) overridden to water (mimic 3D camera cases), and flat-phantom techniques. Radiation dose was prescribed to a depth on the central axis per physician's order, and the monitor units (MUs) were calculated. The 3D camera spatial accuracy was evaluated by comparing the 3D surface of a head phantom captured by a 3D camera and that generated with the CT scan in the treatment planning system. A clinical case is presented, and MUs were calculated using the 3D camera body contour with HU overridden to water.

Results: Across six cases the average change in MUs between the full CT and the 3Dwater (CT scan with HU overridden to water) calculations was 1.3% with a standard deviation of 1.0%. The corresponding hotspots had a mean difference of 0.4% and a standard deviation of 1.9%. The 3D camera captured surface of a head phantom was found to have a 0.59 mm standard deviation from the surface derived from the CT scan. In-vivo dose measurements (213 ± 8 cGy) agreed with the 3D-camera planned dose of 209 ± 6 cGy, compared to 192 ± 6 cGy for the flat-phantom calculation (same MUs).

Conclusions: Electron beam dosimetry is affected by irregular body surfaces. 3D cameras can capture irregular body contours which allow accurate dosimetry of electron beam treatment as an alternative to costly CT scans with no extra exposure to radiation. Tools and workflow for clinical implementation are provided.

Citing Articles

Optical Surface-guided Radiation Therapy for Upper and Lower Limb Sarcomas: An Analysis of Setup Errors and Clinical Target Volume-To-Planning Target Volume Margins.

Zhang Y, Huan F, Wan B, Li Z, Li W, Zhu G Adv Radiat Oncol. 2024; 9(8):101526.

PMID: 39026611 PMC: 11255107. DOI: 10.1016/j.adro.2024.101526.


A Software Suite for Limb Volume Analysis Applicable in Clinical Settings: Upper Limb Quantification.

Nascimben M, Lippi L, Fusco N, Invernizzi M, Rimondini L Front Bioeng Biotechnol. 2023; 10:863689.

PMID: 36798789 PMC: 9928154. DOI: 10.3389/fbioe.2022.863689.


Evaluation of optical 3D scanning system for radiotherapy use.

Crowe S, Luscombe J, Maxwell S, Simpson-Page E, Poroa T, Wilks R J Med Radiat Sci. 2021; 69(2):218-226.

PMID: 34877819 PMC: 9163482. DOI: 10.1002/jmrs.562.

References
1.
Shiu A, Tung S, Hogstrom K, Wong J, Gerber R, Harms W . Verification data for electron beam dose algorithms. Med Phys. 1992; 19(3):623-36. DOI: 10.1118/1.596808. View

2.
Skinner L, Fahimian B, Yu A . Tungsten filled 3D printed field shaping devices for electron beam radiation therapy. PLoS One. 2019; 14(6):e0217757. PMC: 6584017. DOI: 10.1371/journal.pone.0217757. View

3.
Gasparyan S, Ko K, Skinner L, Ko R, Loo Jr B, Fahimian B . Patient motion tracking for non-isocentric and non-coplanar treatments via fixed frame-of-reference 3D camera. J Appl Clin Med Phys. 2020; 21(3):162-166. PMC: 7075370. DOI: 10.1002/acm2.12842. View

4.
Djajaputra D, Li S . Real-time 3D surface-image-guided beam setup in radiotherapy of breast cancer. Med Phys. 2005; 32(1):65-75. DOI: 10.1118/1.1828251. View

5.
Jenkins C, Xing L, Yu A . Using a handheld stereo depth camera to overcome limited field-of-view in simulation imaging for radiation therapy treatment planning. Med Phys. 2017; 44(5):1857-1864. PMC: 5462446. DOI: 10.1002/mp.12207. View