We aimed to assess the accuracy of an artificial intelligence (AI)-based real-time anatomy identification software specifically developed to ease image interpretation intended for ultrasound-guided peripheral nerve block (UGPNB). Forty healthy participants (20 women, 20 men) were enrolled to perform interscalene, supraclavicular, infraclavicular, and transversus abdominis plane (TAP) blocks under ultrasound guidance using AI software by anesthesiology trainees. During block practice by a trainee, once the software indicates 100% scan success of each block associated anatomic landmarks, both raw and labeled ultrasound images were saved, assessed, and validated using a 5-point scale by expert validators. When trainees reached 100% scan success, accuracy scores of the validators were noted. Correlation analysis was used whether the relationship (r) according to demographics (gender, age, and body mass index: BMI) and block type exist. The BMI (kg/m) and age (year) of participants were 22.2 ± 3 and 32.2 ± 5.25, respectively. Assessment scores of validators for all blocks were similar in male and female individuals. Mean assessment scores of validators were not significantly different according to age and BMI except for TAP block, which was inversely correlated with age and BMI (p = 0.01). AI technology can successfully interpret anatomical structures in real-time sonography while assisting young anesthesiologists during UGPNB practice.

Citing Articles

Artificial intelligence in anesthesiology: a bibliometric analysis.

Xie B, Li T, Ma F, Li Q, Xiao Q, Xiong L Perioper Med (Lond). 2024; 13(1):121.

PMID: 39716340 PMC: 11668081. DOI: 10.1186/s13741-024-00480-x.


Optimising artificial intelligence ultrasound tools in anaesthesiology and perioperative medicine: The next frontier for advanced technology application.

Jones A, Tang R, Dabo-Trubelja A, Yeoh C, Richards L, Gottumukkala V Indian J Anaesth. 2024; 68(11):1016-1021.

PMID: 39659528 PMC: 11626875. DOI: 10.4103/ija.ija_578_24.


Current evidence on artificial intelligence in regional anesthesia.

Swain B, Nag D, Anand R, Kumar H, Ganguly P, Singh N World J Clin Cases. 2024; 12(33):6613-6619.

PMID: 39600473 PMC: 11514339. DOI: 10.12998/wjcc.v12.i33.6613.


Variability between human experts and artificial intelligence in identification of anatomical structures by ultrasound in regional anaesthesia: a framework for evaluation of assistive artificial intelligence.

Bowness J, Morse R, Lewis O, Lloyd J, Burckett-St Laurent D, Bellew B Br J Anaesth. 2024; .

PMID: 39492288 PMC: 11103080. DOI: 10.1016/j.bja.2023.09.023.


Artificial intelligence-assisted ultrasound-guided regional anaesthesia: An explorative scoping review.

Marino M, Hagh R, Hamrin Senorski E, Longo U, Oeding J, Nellgard B J Exp Orthop. 2024; 11(3):e12104.

PMID: 39144578 PMC: 11322584. DOI: 10.1002/jeo2.12104.


References
1.
Bowness J, El-Boghdadly K, Burckett-St Laurent D . Artificial intelligence for image interpretation in ultrasound-guided regional anaesthesia. Anaesthesia. 2020; 76(5):602-607. DOI: 10.1111/anae.15212. View

2.
Huang C, Zhou Y, Tan W, Qiu Z, Zhou H, Song Y . Applying deep learning in recognizing the femoral nerve block region on ultrasound images. Ann Transl Med. 2019; 7(18):453. PMC: 6803209. DOI: 10.21037/atm.2019.08.61. View

3.
Jeng C, Torrillo T, Rosenblatt M . Complications of peripheral nerve blocks. Br J Anaesth. 2010; 105 Suppl 1:i97-107. DOI: 10.1093/bja/aeq273. View