Abstract

In this laboratory study, we assessed the resistance to microvacuole (glistening) formation in hydrophobic intraocular lenses (IOLs). Glistenings were induced in five lenses each of five different hydrophobic acrylic IOL models, using an established in vitro laboratory model: 800C (Rayner, Worthing, UK), AcrySof SN60WF (Alcon, Fort Worth, USA), Tecnis ZCB00 (Johnson & Johnson Vision, Santa Ana, USA), Vivinex XY1 (Hoya, Tokyo, Japan) and CT Lucia 611P (Zeiss, Oberkochen, Germany). We evaluated the number of microvacuoles per square millimeter (MV/mm2) in the central part of each IOL. Results were analyzed statistically, and mean glistening numbers were ranked, with the highest in the SN60WF which had 66.0 (±45.5) MVs/mm, followed by the 611P with 30.7 (±8.4) MVs/mm2. The 800C and XY1 showed comparable values of 2.0 (±3.6) and 2.7 (±2.4) MVs/mm2, respectively. ZCB00 had the lowest number with 0.9 (±0.6) MVs/mm2. This study shows that the resistance to glistening formation differs depending on the hydrophobic acrylic copolymer composition of the IOL material. Some IOLs from current clinical use are still prone to develop glistenings whereas others, including the ZCB00, 800C and XY1 show high resistance to microvacuole formation.

Citing Articles

How do intraocular lens materials influence the outcome of cataract surgery?.

Grzybowski A, Auffarth G, LaHood B Curr Opin Ophthalmol. 2024; 36(1):18-24.

PMID: 39446645 PMC: 11620326. DOI: 10.1097/ICU.0000000000001095.


Examining Penetration and Residual Depth in Modern Acrylic Foldable Intraocular Lenses: A Laboratory Study Using Differential Interference Contrast Microscopy to Compare Hydrophilic and Hydrophobic Materials.

Borkenstein A, Borkenstein E Cureus. 2024; 16(9):e70383.

PMID: 39345802 PMC: 11438304. DOI: 10.7759/cureus.70383.


Altering chromatic aberration: how this latest trend in intraocular-lens design affects visual quality in pseudophakic patients.

Labuz G, Gungor H, Auffarth G, Yildirim T, Khoramnia R Eye Vis (Lond). 2023; 10(1):49.

PMID: 38082420 PMC: 10714497. DOI: 10.1186/s40662-023-00367-w.


Early Clinical Experience with a New Hydrophobic Acrylic Single-Piece Monofocal Intraocular Lens.

Schallhorn S, Teenan D, Venter J, Schallhorn J, Hannan S Clin Ophthalmol. 2023; 17:3419-3427.

PMID: 38026603 PMC: 10655610. DOI: 10.2147/OPTH.S433530.


Clinical Outcomes of a Monofocal, Optimized, Aspheric, Hydrophobic Acrylic Intraocular Lens Implant.

Garcia-Tomas B, Marin-Sanchez J, Garcia-Elskamp C, Alcon-Ruiz E, Montesinos-Lopez L, Garcia Martinez-Lozano B Clin Ophthalmol. 2023; 17:3215-3224.

PMID: 37915319 PMC: 10617528. DOI: 10.2147/OPTH.S434378.


References
1.
Ballin N . Glistenings in injection-molded lens. J Am Intraocul Implant Soc. 1984; 10(4):473. DOI: 10.1016/s0146-2776(84)80052-x. View

2.
Khoramnia R, Yildirim T, Labuz G, Mayer C, Auffarth G . [Opacification of intraocular lenses: laboratory and clinical findings]. Ophthalmologe. 2020; 118(7):633-642. PMC: 8260513. DOI: 10.1007/s00347-020-01259-3. View

3.
Miyata K, Honbo M, Otani S, Nejima R, Minami K . Effect on visual acuity of increased surface light scattering in intraocular lenses. J Cataract Refract Surg. 2012; 38(2):221-6. DOI: 10.1016/j.jcrs.2011.08.042. View

4.
Colin J, Praud D, Touboul D, Schweitzer C . Incidence of glistenings with the latest generation of yellow-tinted hydrophobic acrylic intraocular lenses. J Cataract Refract Surg. 2012; 38(7):1140-6. DOI: 10.1016/j.jcrs.2012.01.031. View

5.
Werner L, Thatthamla I, Ong M, Schatz H, Garcia-Gonzalez M, Gros-Otero J . Evaluation of clarity characteristics in a new hydrophobic acrylic IOL in comparison to commercially available IOLs. J Cataract Refract Surg. 2019; 45(10):1490-1497. DOI: 10.1016/j.jcrs.2019.05.017. View