A Multiscale Finite Element Analysis of Balloon Kyphoplasty to Investigate the Risk of Bone-Cement Separation In Vivo

Overview
Date 2021 Apr 26
PMID 33900988
Citations 2
Authors
Affiliations
Soon will be listed here.
Abstract

Background: During the past decade there has been a significant increase in the number of vertebral fractures being treated with the balloon kyphoplasty procedure. Although previous investigations have found kyphoplasty to be an effective treatment for reducing patient pain and lowering cement-leakage risk, there have been reports of vertebral recollapse following the procedure. These reports have indicated evidence of in vivo bone-cement separation leading to collapse of the treated vertebra.

Methods: The following study documents a multiscale analysis capable of evaluating the risk of bone-cement interface separation during lying, standing, and walking activities following balloon kyphoplasty.

Results: Results from the analysis found that instances of reduced cement interlock could initiate both tensile and shear separation of the interface region at up to 7 times the failure threshold during walking or up to 1.9 times the threshold during some cases for standing. Lying prone offered the best protection from interface failure in all cases, with a minimum safety factor of 2.95.

Conclusions: The results of the multiscale analysis show it is essential for kyphoplasty simulations to take account of the micromechanical behavior of the bone-cement interface to be truly representative of the in vivo situation after the treatment. The results further illustrate the importance of ensuring adequate cement infiltration into the compacted bone periphery during kyphoplasty through a combination of new techniques, tools, and biomaterials in a multifaceted approach to solve this complex challenge.

Citing Articles

Adjacent vertebral fractures in the lumbar and thoracic spine after balloon kyphoplasty: A finite element analysis.

Yamaguchi Y, Kobayashi T, Todo M, Hijikata Y, Mawatari M, Tsukamoto M Medicine (Baltimore). 2024; 103(17):e37912.

PMID: 38669425 PMC: 11049699. DOI: 10.1097/MD.0000000000037912.


Bone Cements Used in Vertebral Augmentation: A State-of-the-art Narrative Review.

Williams T, Adler T, Smokoff L, Kaur A, Rodriguez B, Prakash K J Pain Res. 2024; 17:1029-1040.

PMID: 38505504 PMC: 10949389. DOI: 10.2147/JPR.S437827.

References
1.
Ohba T, Ebata S, Clinton D, Koyama K, Haro H . Instability of treated vertebrae after balloon kyphoplasty causing paraparesis in osteoporotic vertebral compression fracture: a report of two cases. Eur Spine J. 2012; 22 Suppl 3:S341-5. PMC: 3641248. DOI: 10.1007/s00586-012-2414-9. View

2.
Janssen D, Mann K, Verdonschot N . Micro-mechanical modeling of the cement-bone interface: the effect of friction, morphology and material properties on the micromechanical response. J Biomech. 2008; 41(15):3158-63. PMC: 2613656. DOI: 10.1016/j.jbiomech.2008.08.020. View

3.
Huiskes R . Failed innovation in total hip replacement. Diagnosis and proposals for a cure. Acta Orthop Scand. 1993; 64(6):699-716. DOI: 10.3109/17453679308994602. View

4.
Tozzi G, Zhang Q, Tong J . 3D real-time micromechanical compressive behaviour of bone-cement interface: experimental and finite element studies. J Biomech. 2011; 45(2):356-63. DOI: 10.1016/j.jbiomech.2011.10.011. View

5.
Polikeit A, Nolte L, Ferguson S . The effect of cement augmentation on the load transfer in an osteoporotic functional spinal unit: finite-element analysis. Spine (Phila Pa 1976). 2003; 28(10):991-6. DOI: 10.1097/01.BRS.0000061987.71624.17. View