Efficacy of Cone-beam Computed Tomography with Modified Gray-scale Range Versus Digital Periapical Radiography for the Assessment of Bone-implant Interface Gaps

Overview
Journal Oral Radiol
Specialty Radiology
Date 2021 Apr 24
PMID 33893899
Citations 1
Authors
Affiliations
Soon will be listed here.
Abstract

Objectives: To compare the efficacy of digital periapical (PA) radiography with cone-beam computed tomography (CBCT) by altering the gray-scale range for the detection of bone-implant interface gaps.

Methods: Titanium implants were placed in 52 blocks of bovine rib such that 42 had no gaps and served as the control group. The implants were subsequently placed in the same blocks with 0.25 mm increase in the osteotomy size to serve as the test group. The remaining 10 blocks were used for random arrangement of the test and control blocks within a fabricated wax arch. CBCT with change in the gray-scale range and digital PA were obtained and evaluated by two observers regarding the presence/absence of gaps. The two imaging modalities were compared by the receiver operating characteristic (ROC) curve, Kappa and McNemar tests (α = 0.05).

Results: PA and CBCT showed moderate agreement for gap detection (k = 0.60). There were no significant differences in the area under the ROC curve (AUC) between CBCT and PA (P = 0.45). The frequency of correct diagnoses on PA radiographs was significantly higher in the test group (P = 0.016), while not significant on the CBCT images (P = 0.344). PA showed higher sensitivity (100%) compared to CBCT (83.33%). However, the specificity of CBCT (92.86%) was greater than that of PA (83.33%).

Conclusions: Certain gray-scale ranges in CBCT enhance the assessment of bone-implant interface which brings the accuracy of CBCT closer to digital PA radiography as the modality of choice.

Citing Articles

Diagnostic accuracy of cone-beam computed tomography with modified grayscale range for detection of buccal cortical plate defects adjacent to dental implants.

Yaghoobi S, Dalili Kajan Z, Khosravifard N, Khalighi Sigaroudi A, Jouybari R Dent Res J (Isfahan). 2023; 20:90.

PMID: 37810449 PMC: 10557990.

References
1.
Branemark P, Hansson B, Adell R, BREINE U, Lindstrom J, HALLEN O . Osseointegrated implants in the treatment of the edentulous jaw. Experience from a 10-year period. Scand J Plast Reconstr Surg Suppl. 1977; 16:1-132. View

2.
Dobbs M, Buckwalter J, SALTZMAN C . Osteoporosis: the increasing role of the orthopaedist. Iowa Orthop J. 2000; 19:43-52. PMC: 1888612. View