A Novel Temperature-Controlled Radiofrequency Catheter Ablation System Used to Treat Patients With Paroxysmal Atrial Fibrillation

Abstract

Objectives: DIAMOND-AF (DiamondTemp™ Ablation System for the Treatment of Paroxysmal Atrial Fibrillation) was a prospective, multicenter, noninferiority, randomized trial that compared the safety and effectiveness of the DTA system versus those of a force-sensing RF ablation system (control) for the treatment of patients with drug-refractory, recurrent, symptomatic paroxysmal atrial fibrillation (AF).

Background: Irrigated radiofrequency (RF) ablation catheters lose tissue temperature acuity, which is vital in assessing lesion formation. DiamondTemp Ablation (DTA) was designed to re-establish accurate tissue temperature measurements during ablation.

Methods: A total of 482 patients with paroxysmal AF were randomized (239 DTA, 243 control) to undergo pulmonary vein isolation and were followed up at 23 sites. Patients were screened for disease progression, cardiac characteristics, and prior interventions. Primary endpoints were effectiveness (freedom from atrial arrhythmia recurrence) and safety (composite of procedure- and device-related serious adverse events).

Results: The primary safety event rate was 3.3% in the DTA group versus 6.6% in the control group (p < 0.001 vs. 6.5% noninferiority margin). Primary effectiveness was met in 79.1% of DTA subjects and 75.7% of control subjects (p < 0.001 vs. -12.5% noninferiority margin). Secondary endpoint analysis found that off-drug effectiveness favored DTA compared with the control (142 [59.4%] vs. 120 [49.4%], respectively; p = 0.03). Total RF time and individual RF ablation duration were significantly shorter with less saline infused through the DTA catheter (p < 0.001). Both arms saw clinically meaningful improvements in quality of life at 12 months.

Conclusions: Safety and efficacy of the DTA system proved noninferior to force-sensing RF ablation in a paroxysmal AF population. Efficiencies were observed using DTA with shorter total RF times, individual RF ablation durations, and less saline infusion. (DiamondTemp™ Ablation System for the Treatment of Paroxysmal Atrial Fibrillation; NCT03334630).

Citing Articles

Temperature-controlled Ablation Versus Conventional Ablation for Pulmonary Vein Isolation in the Treatment of AF: A Systematic Review and Meta-Analysis.

Clay B, Sokhal B, Zeriouh S, Srinivasan N, Gajendragadkar P, Martin C Arrhythm Electrophysiol Rev. 2025; 14:e03.

PMID: 40017671 PMC: 11865669. DOI: 10.15420/aer.2024.41.


Comparison of a Novel Temperature-Controlled Diamond-Tip Catheter and a Power-Controlled Gold-Tip Catheter for the Irrigated Ablation of Cavotricuspid Isthmus-Dependent Atrial Flutter.

Auf der Maur E, Kueffer T, Thalmann G, Kozhuharov N, Galuszka O, Iqbal S J Clin Med. 2025; 14(3).

PMID: 39941371 PMC: 11818538. DOI: 10.3390/jcm14030701.


2024 European Heart Rhythm Association/Heart Rhythm Society/Asia Pacific Heart Rhythm Society/Latin American Heart Rhythm Society expert consensus statement on catheter and surgical ablation of atrial fibrillation.

Tzeis S, Gerstenfeld E, Kalman J, Saad E, Sepehri Shamloo A, Andrade J J Arrhythm. 2024; 40(6):1217-1354.

PMID: 39669937 PMC: 11632303. DOI: 10.1002/joa3.13082.


High-Power Short-Duration Ablation of Paroxysmal and Persistent Atrial Fibrillation.

Hoyos C, Matos C, Miranda-Arboleda A, Patino C, Hincapie D, Osorio J Rev Cardiovasc Med. 2024; 24(12):363.

PMID: 39077093 PMC: 11272868. DOI: 10.31083/j.rcm2412363.


2024 European Heart Rhythm Association/Heart Rhythm Society/Asia Pacific Heart Rhythm Society/Latin American Heart Rhythm Society expert consensus statement on catheter and surgical ablation of atrial fibrillation.

Tzeis S, Gerstenfeld E, Kalman J, Saad E, Sepehri Shamloo A, Andrade J J Interv Card Electrophysiol. 2024; 67(5):921-1072.

PMID: 38609733 DOI: 10.1007/s10840-024-01771-5.