» Articles » PMID: 33503117

VALIDATION OF A NEW WATER-PERFUSED HIGH-RESOLUTION MANOMETRY SYSTEM

Overview
Date 2021 Jan 27
PMID 33503117
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Water-perfused high resolution manometry system. High-resolution manometry is more costly but clinically superior to conventional manometry. Water-perfused systems may decrease costs, but it is unclear if they are as reliable as solid-state systems, and reference values are interchangeable.

Aim: To validate normal values for a new water-perfusion high-resolution manometry system.

Methods: Normative values for a 24-sensors water perfused high-resolution manometry system were validated by studying 225 individuals who underwent high resolution manometry for clinical complaints. Patients were divided in four groups: group 1 - gastroesophageal reflux disease; group 2 - achalasia; group 3 - systemic diseases with possible esophageal manifestation; and group 4 - dysphagia.

Results: In group 1, a hypotonic lower esophageal sphincter was found in 49% of individuals with positive 24 h pH monitoring, and in 28% in pH-negative individuals. In groups 2 and 3, aperistalsis was found in all individuals. In group 4, only one patient (14%) had normal high-resolution manometry.

Conclusions: The normal values determined for this low-cost water-perfused HRM system with unique peristaltic pump and helicoidal sensor distribution are discriminatory of most abnormalities of esophageal motility seen in clinical practice.

Citing Articles

MANAGEMENT OF SYMPTOMS RECURRENCE AFTER MYOTOMY FOR ACHALASIA. A PRACTICAL APPROACH.

Tustumi F, Szachnowicz S, Andreollo N, Seguro F, Bianchi E, Duarte A Arq Bras Cir Dig. 2023; 36:e1780.

PMID: 38088726 PMC: 10712922. DOI: 10.1590/0102-672020230062e1780.


Endoluminal Procedures and Devices for Esophageal Tract Investigation: A Critical Review.

Spreafico G, Chiurazzi M, Bagnoli D, Emiliani S, De Bortoli N, Ciuti G Sensors (Basel). 2023; 23(21).

PMID: 37960557 PMC: 10650290. DOI: 10.3390/s23218858.


LAPAROSCOPIC ANTIREFLUX SURGERY: WERE OLD QUESTIONS ANSWERED? PARTIAL OR TOTAL FUNDOPLICATION?.

Allaix M, Rebecchi F, Bellocchia A, Morino M, Patti M Arq Bras Cir Dig. 2023; 36:e1741.

PMID: 37436210 PMC: 10335357. DOI: 10.1590/0102-672020230023e1741.


LAPAROSCOPIC ANTIREFLUX SURGERY: ARE OLD QUESTIONS ANSWERED? SHOULD IT BE USED CONJOINED WITH ENDOSCOPIC THERAPY FOR BARRETT'S ESOPHAGUS?.

Han S, Low D Arq Bras Cir Dig. 2022; 35:e1664.

PMID: 35766609 PMC: 9254380. DOI: 10.1590/0102-672020210002e1664.


THE DYSPHAGIA INVESTIGATION: IS THERE STILL SPACE FOR THE VIDEOFLUOROSCOPIC METHOD?.

Marques C, Abrahao-Junior L, Lemme E Arq Bras Cir Dig. 2022; 35:e1650.

PMID: 35730879 PMC: 9254383. DOI: 10.1590/0102-672020210002e1650.


References
1.
Kahrilas P, Bredenoord A, Fox M, Gyawali C, Roman S, Smout A . The Chicago Classification of esophageal motility disorders, v3.0. Neurogastroenterol Motil. 2014; 27(2):160-74. PMC: 4308501. DOI: 10.1111/nmo.12477. View

2.
Reddy C, Baker J, Lau J, Chen J . High-Resolution Manometry Diagnosis of Ineffective Esophageal Motility Is Associated with Higher Reflux Burden. Dig Dis Sci. 2019; 64(8):2199-2205. DOI: 10.1007/s10620-019-05633-3. View

3.
Tseng P, Wong R, Wu J, Chen C, Tu C, Lee Y . Normative values and factors affecting water-perfused esophageal high-resolution impedance manometry for a Chinese population. Neurogastroenterol Motil. 2017; 30(6):e13265. DOI: 10.1111/nmo.13265. View

4.
Vicentine F, Herbella F, Allaix M, Silva L, Patti M . Comparison of idiopathic achalasia and Chagas' disease esophagopathy at the light of high-resolution manometry. Dis Esophagus. 2013; 27(2):128-33. DOI: 10.1111/dote.12098. View

5.
Triadafilopoulos G, Tandon A, Shetler K, Clarke J . Clinical and pH study characteristics in reflux patients with and without ineffective oesophageal motility (IEM). BMJ Open Gastroenterol. 2017; 3(1):e000126. PMC: 5174815. DOI: 10.1136/bmjgast-2016-000126. View