The Reconstruction of a 12-Lead Electrocardiogram from a Reduced Lead Set Using a Focus Time-Delay Neural Network

Overview
Date 2021 Jan 25
PMID 33488027
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

Background: The 12-lead electrocardiogram (ECG) is the gold-standard ECG method used by cardiologists. However, accurate electrode placement is difficult and time consuming, and can lead to incorrect interpretation.

Objectives: The objective of this study was to accurately reconstruct a full 12-lead ECG from a reduced lead set.

Methods: Five-electrode placement was used to generate leads I, II, III, aVL, aVR, aVF and V2. These seven leads served as inputs to the focus time-delay neural network (FTDNN) which derived the remaining five precordial leads (V1, V3-V6). An online archived medical database containing 549 cases of ECG recordings was used to train, validate and test the FTDNN.

Results: After removing outliers, the reconstructed leads exhibited correlation values of between 0.8609 and 0.9678 as well as low root mean square error values of between 123 μV and 245 μV across all cases, for both healthy controls and cardiovascular disease subgroups except the bundle branch block disease subgroup. The results of the FTDNN method compared favourably to those of prior lead reconstruction methods.

Conclusions: A standard 12-lead ECG was successfully reconstructed with high quantitative correlations from a reduced lead set using only five electrodes, of which four were placed on the limbs. Less reliance on precordial leads will aid in the reduction of electrode placement errors, ultimately improving ECG lead accuracy and reduce the number of cases that are incorrectly diagnosed.

Citing Articles

A novel method for 12-lead ECG reconstruction.

EPMoghaddam D, Banta A, Post A, Razavi M, Aazhang B Conf Rec Asilomar Conf Signals Syst Comput. 2024; 2023:1054-1058.

PMID: 39286539 PMC: 11404295. DOI: 10.1109/ieeeconf59524.2023.10476822.


AI-enhanced reconstruction of the 12-lead electrocardiogram via 3-leads with accurate clinical assessment.

Mason F, Pandey A, Gadaleta M, Topol E, Muse E, Quer G NPJ Digit Med. 2024; 7(1):201.

PMID: 39090394 PMC: 11294561. DOI: 10.1038/s41746-024-01193-7.


Wearable 12-Lead ECG Acquisition Using a Novel Deep Learning Approach from Frank or EASI Leads with Clinical Validation.

Fu F, Zhong D, Liu J, Xu T, Shen Q, Wang W Bioengineering (Basel). 2024; 11(3).

PMID: 38534567 PMC: 10967908. DOI: 10.3390/bioengineering11030293.


12-Lead ECG Reconstruction Based on Data From the First Limb Lead.

Savostin A, Koshekov K, Ritter Y, Savostina G, Ritter D Cardiovasc Eng Technol. 2024; 15(3):346-358.

PMID: 38424391 DOI: 10.1007/s13239-024-00719-0.


Design and technical validation to generate a synthetic 12-lead electrocardiogram dataset to promote artificial intelligence research.

Yoo H, Moon J, Kim J, Joo H Health Inf Sci Syst. 2023; 11(1):41.

PMID: 37662618 PMC: 10468461. DOI: 10.1007/s13755-023-00241-y.


References
1.
Nelwan S, Kors J, Crater S, Meij S, van Dam T, Simoons M . Simultaneous comparison of 3 derived 12-lead electrocardiograms with standard electrocardiogram at rest and during percutaneous coronary occlusion. J Electrocardiol. 2008; 41(3):230-7. DOI: 10.1016/j.jelectrocard.2008.01.011. View

2.
Rudiger A, Hellermann J, Mukherjee R, Follath F, Turina J . Electrocardiographic artifacts due to electrode misplacement and their frequency in different clinical settings. Am J Emerg Med. 2007; 25(2):174-8. DOI: 10.1016/j.ajem.2006.06.018. View

3.
Lo C, Chang S, Tsai J, Kuo J, Chen Y, Huang M . Evaluation of the Accuracy of ECG Captured by CardioChip through Comparison of Lead I Recording to a Standard 12-Lead ECG Recording Device. Acta Cardiol Sin. 2018; 34(2):144-151. PMC: 5863068. DOI: 10.6515/ACS.201803_34(2).20170919A. View

4.
Tsouri G, Ostertag M . Patient-specific 12-lead ECG reconstruction from sparse electrodes using independent component analysis. IEEE J Biomed Health Inform. 2014; 18(2):476-82. DOI: 10.1109/JBHI.2013.2294561. View

5.
KERWIN A, McLean R, TEGELAAR H . A method for the accurate placement of chest electrodes in the taking of serial electrocardiographic tracings. Can Med Assoc J. 1960; 82:258-61. PMC: 1937716. View