Development and Accuracy Evaluation of a Single-camera Intra-bore Surface Scanning System for Radiotherapy in an O-ring Linac

Overview
Specialty Oncology
Date 2021 Jan 18
PMID 33458272
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

Background And Purpose: Current commercial surface scanning systems are not able to monitor patients during radiotherapy fractions in closed-bore linacs during adaptive workflows. In this work a surface scanning system for monitoring in an O-ring linac is proposed.

Methods And Materials: A depth camera was mounted at the backend of the bore. The acquired surface point cloud was transformed to the linac coordinate system after a cube detection calibration step. The real-time surface was registered using an Iterative Closest Point algorithm to a reference region-of-interest of the body contour from the planning CT and of a depth camera surface acquisition from the first fraction. The positioning accuracy was investigated using anthropomorphic 3D-printed phantoms with embedded markers: a head, hand and breast. To simulate clinically observed positioning errors, each phantom was placed 24 times with 0-10 mm and 0-8° offsets from the planned position. At every position a cone-beam CT (CBCT) was acquired and a surface registration performed. The surface registration error was determined as the difference between the surface registration and the CBCT-to-CT fiducial marker registration.

Results: The registration errors were (mean ± SD): lat: 0.4 ± 0.8 mm, vert: -0.2 ± 0.2 mm, long: 0.3 ± 0.5 mm and Yaw: -0.2 ± 0.6°, Pitch: 0.4 ± 0.2°, Roll: 0.5 ± 0.8° for the body contour reference, and lat: -0.7 ± 0.7 mm, vert: 0.3 ± 0.2 mm, long: 0.2 ± 0.5 mm and Yaw: -0.5 ± 0.5°, Pitch: 0.1 ± 0.3°, Roll: -0.7 ± 0.7° for the captured surface reference.

Conclusion: The proposed single camera intra-bore surface system was capable of accurately detecting phantom displacements and allows intrafraction motion monitoring for surface guided radiotherapy inside the bore of O-ring gantries.

Citing Articles

Surface guided ring gantry radiotherapy in deep inspiration breath hold for breast cancer patients.

Kadhim M, Haraldsson A, Kugele M, Enocson H, Back S, Ceberg S J Appl Clin Med Phys. 2024; 25(10):e14463.

PMID: 39138877 PMC: 11466460. DOI: 10.1002/acm2.14463.


Reproducibility of surface-based deep inspiration breath-hold technique for lung stereotactic body radiotherapy on a closed-bore gantry linac.

Nguyen D, Reinoso R, Farah J, Yossi S, Lorchel F, Passerat V Phys Imaging Radiat Oncol. 2023; 26:100448.

PMID: 37252251 PMC: 10213090. DOI: 10.1016/j.phro.2023.100448.


A method for beam's eye view breath-hold monitoring during breast volumetric modulated arc therapy.

Carr M, Gargett M, Stanton C, Zwan B, Byrne H, Booth J Phys Imaging Radiat Oncol. 2023; 25:100419.

PMID: 36875326 PMC: 9975298. DOI: 10.1016/j.phro.2023.100419.


Advances and potential of optical surface imaging in radiotherapy.

Li G Phys Med Biol. 2022; 67(16).

PMID: 35868290 PMC: 10958463. DOI: 10.1088/1361-6560/ac838f.


Accuracy evaluation of surface registration algorithm using normal distribution transform in stereotactic body radiotherapy/radiosurgery: A phantom study.

Lee H, Park J, Kim K, Lee D, Sohn M J Appl Clin Med Phys. 2022; 23(3):e13521.

PMID: 34985179 PMC: 8906233. DOI: 10.1002/acm2.13521.


References
1.
Noonan P, Howard J, Hallett W, Gunn R . Repurposing the Microsoft Kinect for Windows v2 for external head motion tracking for brain PET. Phys Med Biol. 2015; 60(22):8753-66. DOI: 10.1088/0031-9155/60/22/8753. View

2.
Gaisberger C, Steininger P, Mitterlechner B, Huber S, Weichenberger H, Sedlmayer F . Three-dimensional surface scanning for accurate patient positioning and monitoring during breast cancer radiotherapy. Strahlenther Onkol. 2013; 189(10):887-93. DOI: 10.1007/s00066-013-0358-6. View

3.
Schonecker S, Walter F, Freislederer P, Marisch C, Scheithauer H, Harbeck N . Treatment planning and evaluation of gated radiotherapy in left-sided breast cancer patients using the Catalyst/Sentinel system for deep inspiration breath-hold (DIBH). Radiat Oncol. 2016; 11(1):143. PMC: 5080745. DOI: 10.1186/s13014-016-0716-5. View

4.
Placht S, Stancanello J, Schaller C, Balda M, Angelopoulou E . Fast time-of-flight camera based surface registration for radiotherapy patient positioning. Med Phys. 2012; 39(1):4-17. DOI: 10.1118/1.3664006. View

5.
Bert C, Metheany K, Doppke K, Taghian A, Powell S, Chen G . Clinical experience with a 3D surface patient setup system for alignment of partial-breast irradiation patients. Int J Radiat Oncol Biol Phys. 2006; 64(4):1265-74. DOI: 10.1016/j.ijrobp.2005.11.008. View