» Articles » PMID: 33213814

1-Year Prospective Evaluation of Clinical Outcomes and Shocks: The Subcutaneous ICD Post Approval Study

Overview
Date 2020 Nov 20
PMID 33213814
Citations 18
Authors
Affiliations
Soon will be listed here.
Abstract

Objectives: This study evaluated spontaneous arrhythmias and clinical outcomes in the S-ICD System PAS (Subcutaneous Implantable Cardioverter-Defibrillator Post Approval Study) cohort.

Background: The U.S. S-ICD PAS trial patient population more closely resembles transvenous ICD cohorts than earlier studies, which included many patients with little structural heart disease and few comorbidities. Early outcomes and low peri-operative complication rates were demonstrated in the S-ICD PAS cohort, but there are no data detailing spontaneous arrhythmias and clinical outcomes.

Methods: The S-ICD PAS prospective registry included 1,637 de novo patients from 86 U.S. centers. Descriptive statistics, Kaplan-Meier time to event, and multivariate logistic regression were performed using data out to 365 days.

Results: Patients (68.5% men; mean ejection fraction of 32.0%; 42.9% ischemic; 13.4% on dialysis) underwent implantation for primary (76.6%) or secondary prevention indication. The complication-free rate was 92.5%. The appropriate shock (AS) rate was 5.3%. A total of 395 ventricular tachycardia (VT) or fibrillation (VF) episodes were appropriately sensed, with 131 (33.2%) self-terminating. First and final shock efficacy (up to 5 shocks) for the 127 discrete AS episodes were 91.3% and 100.0%, respectively. Discrete AS episodes included 67 monomorphic VT (MVT) and 60 polymorphic VT (PVT)/VF, with first shock efficacy of 95.2% and 86.7%, respectively. There were 19 storm events in 18 subjects, with 84.2% conversion success. Storm episodes were more likely PVT/VF (98 of 137).

Conclusions: In the first year after implantation, a predominantly primary prevention population with low ejection fraction demonstrated a high complication-free rate and spontaneous event shock efficacy for MVT and PVT/VF arrhythmias at rapid ventricular rates. (Subcutaneous Implantable Cardioverter-Defibrillator System Post Approval Study [S-ICD PAS; NCT01736618).

Citing Articles

The Role of Subcutaneous ICDs in the Prevention of Sudden Cardiac Death.

John L, Karimianpour A, Gold M US Cardiol. 2024; 15:e19.

PMID: 39720500 PMC: 11664769. DOI: 10.15420/usc.2021.01.


Performance of Subcutaneous Implantable Cardioverter Defibrillator (S-ICD) in Chinese Population with Primary Prevention Indications: A Prospective Observational Cohort Study.

Li Y, Chen Y, Wang J, Xu J, Li R, Qiu Z Med Sci Monit. 2024; 30:e942747.

PMID: 38400538 PMC: 10900845. DOI: 10.12659/MSM.942747.


Inappropriate Subcutaneous Implantable Cardioverter-defibrillator Shocks-A Rare Case of Triple Counting.

Ghanta S, Alotaibi B, Paydak H, Mounsey J, Vallurupalli S, Devabhaktuni S J Innov Card Rhythm Manag. 2023; 14(12):5670-5674.

PMID: 38155720 PMC: 10752427. DOI: 10.19102/icrm.2023.14121.


Subcutaneous Implantable Cardioverter Defibrillator: A Contemporary Overview.

Guarracini F, Preda A, Bonvicini E, Coser A, Martin M, Quintarelli S Life (Basel). 2023; 13(8).

PMID: 37629509 PMC: 10455445. DOI: 10.3390/life13081652.


Real-world evidence for the use of subcutaneous implantable cardioverter-defibrillators in China: A single-center experience.

Zhang L, Li X, Liang Y, Wang J, Li M, Pan L Herz. 2023; 48(6):462-469.

PMID: 37540305 DOI: 10.1007/s00059-023-05192-4.