Suitable Hemolysis Index for Low-flow Rotary Blood Pumps
Overview
Affiliations
A suitable index is needed for hemolysis tests that use low-flow pumps, such as pediatric blood pumps or blood purification pumps. To create such an index, the present study investigates the change of plasma-free hemoglobin in the pump circuit with time and the change of the hemolysis rate with flow rate and impeller rotational speed. The results show that the hemolysis rate or the increase rate of the total free hemoglobin are suitable measures for hemolysis evaluation for low-flow pumps.
Axial-flow polymer bridge pump with hydrodynamic bearings.
Yamane T, Akao E, Kashiwazaki T, Koda Y, Tanaka H J Artif Organs. 2024; .
PMID: 39277834 DOI: 10.1007/s10047-024-01465-9.
The design and evaluation of the outflow structures of an interventional microaxial blood pump.
Yun Z, Yao J, Wang L, Tang X, Feng Y Front Physiol. 2023; 14:1169905.
PMID: 37250127 PMC: 10213901. DOI: 10.3389/fphys.2023.1169905.