Noise Detection in Electrocardiogram Signals for Intensive Care Unit Patients

Overview
Journal IEEE Access
Date 2020 Nov 2
PMID 33133877
Citations 19
Authors
Affiliations
Soon will be listed here.
Abstract

Long term electrocardiogram (ECG) signals recorded in an intensive care unit (ICU) are often corrupted by severe motion and noise artifacts (MNA), which may lead to many false alarms including inaccurate detection of atrial fibrillation (AF). We developed an automated method to detect MNA from ECG recordings in the Medical Information Mart for Intensive Care (MIMIC) III database. Since AF detection is often based on identification of irregular RR intervals derived from the QRS complexes, the main design focus of our MNA detection algorithm was to identify corrupted QRS complexes of the ECG signals. The MNA in the MIMIC III database contain not only motion-induced noise, but also a plethora of non-ECG waveforms, which must also be automatically identified. Our algorithm is designed to first discriminate between ECG and non-ECG waveforms using both time and spectral-domain properties. For the segments of data containing ECG waveforms, a time-frequency spectrum and its sub-band decomposition approach were used to identify MNA, and high frequency noise ECG segments, respectively. The algorithm was tested on data from 35 subjects in normal sinus rhythm and 25 AF subjects. The proposed method is shown to accurately discriminate between segments that contained real ECG waveforms and those that did not, even though the latter were numerous in some subjects. In addition, we found a significant reduction (> 94%) in false positive detection of AF in normal subjects when our MNA detection algorithm was used. Without using it, we inaccurately detected AF owing to the MNA.

Citing Articles

The MSPTDfast photoplethysmography beat detection algorithm: design, benchmarking, and open-source distribution.

Charlton P, Arguello-Prada E, Arguello Prada E, Mant J, Kyriacou P, Kyriacou P Physiol Meas. 2025; 46(3).

PMID: 39978069 PMC: 11894679. DOI: 10.1088/1361-6579/adb89e.


LieRHRV system for remote lie detection using heart rate variability parameters.

Davoodi M, Aspis N, Drori Y, Weiser-Bitoun I, Yaniv Y Sci Rep. 2024; 14(1):30749.

PMID: 39730487 PMC: 11681120. DOI: 10.1038/s41598-024-80480-5.


Compressed Deep Learning Models for Wearable Atrial Fibrillation Detection through Attention.

Makynen M, Ng G, Li X, Schlindwein F, Pearce T Sensors (Basel). 2024; 24(15).

PMID: 39123835 PMC: 11314646. DOI: 10.3390/s24154787.


Performance comparison of 6 in-hospital patient monitoring systems in the detection and alarm of ventricular cardiac arrhythmias.

Cosentino N, Zhang X, Farrar E, Yapici H, Coffeng R, Vaananen H Cardiovasc Digit Health J. 2024; 5(2):70-77.

PMID: 38765622 PMC: 11096657. DOI: 10.1016/j.cvdhj.2024.02.002.


Wearable Sensors as a Preoperative Assessment Tool: A Review.

Syversen A, Dosis A, Jayne D, Zhang Z Sensors (Basel). 2024; 24(2).

PMID: 38257579 PMC: 10820534. DOI: 10.3390/s24020482.


References
1.
Mateo J, Laguna P . Analysis of heart rate variability in the presence of ectopic beats using the heart timing signal. IEEE Trans Biomed Eng. 2003; 50(3):334-43. DOI: 10.1109/TBME.2003.808831. View

2.
Orphanidou C, Bonnici T, Charlton P, Clifton D, Vallance D, Tarassenko L . Signal-quality indices for the electrocardiogram and photoplethysmogram: derivation and applications to wireless monitoring. IEEE J Biomed Health Inform. 2014; 19(3):832-8. DOI: 10.1109/JBHI.2014.2338351. View

3.
Chon K, Dash S, Ju K . Estimation of respiratory rate from photoplethysmogram data using time-frequency spectral estimation. IEEE Trans Biomed Eng. 2009; 56(8):2054-63. DOI: 10.1109/TBME.2009.2019766. View

4.
Wang H, Siu K, Ju K, Chon K . A high resolution approach to estimating time-frequency spectra and their amplitudes. Ann Biomed Eng. 2006; 34(2):326-38. DOI: 10.1007/s10439-005-9035-y. View

5.
Dash S, Chon K, Lu S, Raeder E . Automatic real time detection of atrial fibrillation. Ann Biomed Eng. 2009; 37(9):1701-9. DOI: 10.1007/s10439-009-9740-z. View