Efficient Hybrid 3D System Calibration for Magnetic particle imaging Systems Using a Dedicated Device

Overview
Journal Sci Rep
Specialty Science
Date 2020 Oct 29
PMID 33116183
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

Image reconstruction in magnetic particle imaging is often performed using a system matrix based approach. The acquisition of a system matrix is a time-consuming calibration which may take several weeks and thus, is not feasible for a clinical device. Due to hardware characteristics of the receive chain, a system matrix may not even be used in similar devices but has to be acquired for each imager. In this work, a dedicated device is used for measuring a hybrid system matrix. It is shown that the measurement time of a 3D system matrix is reduced by 96%. The transfer function of the receive chains is measured, which allows the use of the same system matrix in multiple devices. Equivalent image reconstruction results are reached using the hybrid system matrix. Furthermore, the inhomogeneous sensitivity profile of receive coils is successfully applied to a hybrid system matrix. It is shown that each aspect of signal acquisition in magnetic particle imaging can be taken into account using hybrid system matrices. It is favourable to use a hybrid system matrix for image reconstruction in terms of measurement time, signal-to-noise ratio and discretisation.

Citing Articles

Open-source device for high sensitivity magnetic particle spectroscopy, relaxometry, and hysteresis loop tracing.

Mattingly E, Barksdale A, Sliwiak M, Chacon-Caldera J, Mason E, Wald L Rev Sci Instrum. 2024; 95(6).

PMID: 38921057 PMC: 11210977. DOI: 10.1063/5.0191946.


Machine Learning and Deep Learning Applications in Magnetic Particle Imaging.

Nigam S, Gjelaj E, Wang R, Wei G, Wang P J Magn Reson Imaging. 2024; 61(1):42-51.

PMID: 38358090 PMC: 11324856. DOI: 10.1002/jmri.29294.


The Reconstruction of Magnetic Particle Imaging: Current Approaches Based on the System Matrix.

Chen X, Jiang Z, Han X, Wang X, Tang X Diagnostics (Basel). 2021; 11(5).

PMID: 33925830 PMC: 8146641. DOI: 10.3390/diagnostics11050773.

References
1.
Graeser M, von Gladiss A, Weber M, M Buzug T . Two dimensional magnetic particle spectrometry. Phys Med Biol. 2017; 62(9):3378-3391. DOI: 10.1088/1361-6560/aa5bcd. View

2.
Ferguson R, Khandhar A, Kemp S, Arami H, Saritas E, Croft L . Magnetic particle imaging with tailored iron oxide nanoparticle tracers. IEEE Trans Med Imaging. 2014; 34(5):1077-84. PMC: 4428902. DOI: 10.1109/TMI.2014.2375065. View

3.
Knopp T, Biederer S, Sattel T, Weizenecker J, Gleich B, Borgert J . Trajectory analysis for magnetic particle imaging. Phys Med Biol. 2008; 54(2):385-97. DOI: 10.1088/0031-9155/54/2/014. View

4.
Knopp T, Biederer S, Sattel T, Rahmer J, Weizenecker J, Gleich B . 2D model-based reconstruction for magnetic particle imaging. Med Phys. 2010; 37(2):485-91. DOI: 10.1118/1.3271258. View

5.
Gleich B, Weizenecker J . Tomographic imaging using the nonlinear response of magnetic particles. Nature. 2005; 435(7046):1214-7. DOI: 10.1038/nature03808. View