Inter- and Intra-unit Reliability of the COSMED K5: Implications for Multicentric and Longitudinal Testing

Overview
Journal PLoS One
Date 2020 Oct 23
PMID 33096546
Citations 1
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To evaluate the intra-unit (RELINTRA) and inter-unit reliability (RELINTER) of two structurally identical units of the metabolic analyser K5 (COSMED, Rome, Italy) that allows to utilize either breath-by-breath (BBB) or dynamic mixing chamber (DMC) technology.

Methods: Identical flow- and gas-signals were transmitted to both K5s that always operated simultaneously either in BBB- or DMC-mode. To assess RELINTRA and RELINTER, a metabolic simulator was applied to simulate four graded levels of respiration. RELINTRA and RELINTER were expressed as typical error (TE%) and Intraclass Correlation Coefficient (ICC). To assess also inter-unit differences via natural respiratory signals, 12 male athletes performed one incremental bike step test each in BBB- and DMC-mode. Inter-unit differences within biological testing were expressed as percentages.

Results: In BBB, TE% of RELINTRA ranged 0.30-0.67 vs. RELINTER 0.16-1.39 and ICC ranged 0.57-1.00 vs. 0.09-1.00. In DMC, TE% of RELINTRA ranged 0.38-0.90 vs. RELINTER 0.03-0.86 and ICC ranged 0.22-1.00 vs. 0.52-1.00. Mean inter-unit differences ranged -2.30-2.20% (Cohen's ds (ds) 0.13-1.52) for BBB- and -0.55-0.61% (ds 0.00-0.65) for DMC-mode, respectively. Inter-unit differences for [Formula: see text] and RER were significant (p < 0.05) at each step.

Conclusion: Two structurally identical K5-units demonstrated accurate RELINTRA with TE < 2.0% and similar RELINTER during metabolic simulation. During biological testing, inter-unit differences for [Formula: see text] and RER in BBB-mode were higher than 2% with partially large ES in BBB. Hence, the K5 should be allocated personally wherever possible. Otherwise, e.g. in multicenter studies, a decrease in total reliability needs to be considered especially when the BBB-mode is applied.

Citing Articles

Effects of 16 weeks of pyramidal and polarized training intensity distributions in well-trained endurance runners.

Filipas L, Bonato M, Gallo G, Codella R Scand J Med Sci Sports. 2021; 32(3):498-511.

PMID: 34792817 PMC: 9299127. DOI: 10.1111/sms.14101.

References
1.
Rosdahl H, Lindberg T, Edin F, Nilsson J . The Moxus Modular metabolic system evaluated with two sensors for ventilation against the Douglas bag method. Eur J Appl Physiol. 2012; 113(5):1353-67. DOI: 10.1007/s00421-012-2551-1. View

2.
Hopkins W . Measures of reliability in sports medicine and science. Sports Med. 2000; 30(1):1-15. DOI: 10.2165/00007256-200030010-00001. View

3.
Saunders P, Telford R, Pyne D, Gore C, Hahn A . Improved race performance in elite middle-distance runners after cumulative altitude exposure. Int J Sports Physiol Perform. 2009; 4(1):134-8. DOI: 10.1123/ijspp.4.1.134. View

4.
Clark S, Bourdon P, Schmidt W, Singh B, Cable G, Onus K . The effect of acute simulated moderate altitude on power, performance and pacing strategies in well-trained cyclists. Eur J Appl Physiol. 2007; 102(1):45-55. DOI: 10.1007/s00421-007-0554-0. View

5.
Gore C, Catcheside P, French S, Bennett J, LaForgia J . Automated VO2max calibrator for open-circuit indirect calorimetry systems. Med Sci Sports Exerc. 1997; 29(8):1095-103. DOI: 10.1097/00005768-199708000-00016. View