Potential of Contrast Agents Based on High-Z Elements for Contrast-enhanced Photon-counting Computed Tomography

Overview
Journal Med Phys
Specialty Biophysics
Date 2020 Oct 4
PMID 33011992
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: In clinics, only iodine- and barium-based contrast agents are currently used for contrast-enhanced x-ray computed tomography (CT). Recently, the introduction of new photon-counting (PC) detectors increased the interest in developing new contrast agents based on heavier elements. These elements may provide more contrast and spectral information compared to iodine and barium thanks to their k-edges at higher energies. In this paper, the potential of high-Z elements in contrast-enhanced CT was evaluated for different patient sizes and x-ray spectra using a PC detector.

Methods: An adult liver phantom with five high-Z element solutions (iodine, gadolinium, ytterbium, tungsten, and bismuth) was scanned with a whole-body photon-counting computed tomography (PCCT) prototype. For each element, the contrast-to-noise ratio at unit concentration and at unit dose (CNRCD) was evaluated in low threshold images ( ) as function of the tube voltage (80, 100, 120, and 140 kV) and in bin images (tube voltage = 120 kV) as function of the higher threshold ( and ). Simulations were performed for validation with measurements and to investigate more elements (cerium and gold), different patient sizes (infant, adult, and obese), and spectrum filtration (with and without 0.4-mm tin filter). The dose reductions associated with the CNRCD improvements over iodine were quantified as well.

Results: CNRCD improvements and dose reductions depend on the investigated scenario. For the infant phantom, dose reductions around 30% were reached using cerium or gadolinium in combination with the tin filter. For the adult and obese phantom, reductions around 50% were provided by gadolinium or ytterbium in combination with the tin filter. Independently of the high-Z element, the CNRCD of two optimally combined bin images was higher than the CNRCD of the low threshold image. Good agreement was found between measurements and simulations.

Conclusions: Between the investigated elements, gadolinium resulted to have the highest potential as novel contrast agent in PCCT, providing significant dose reductions for all patient sizes. Compared to the other elements, the implementation of gadolinium as CT contrast agent may be facilitated since it is already deployed as contrast agents for magnetic resonance imaging.

Citing Articles

Nanoparticle Contrast Agents for Photon-Counting Computed Tomography: Recent Developments and Future Opportunities.

Devkota L, Bhavane R, Badea C, Tanifum E, Annapragada A, Ghaghada K Wiley Interdiscip Rev Nanomed Nanobiotechnol. 2025; 17(1):e70004.

PMID: 39948059 PMC: 11874078. DOI: 10.1002/wnan.70004.


Composite Contrast Enhancement of Hydrogel-Based Implants for Photon-Counting Computed Tomography Studies.

Suslova E, Shashurin D, Maslakov K, Kupreenko S, Luneva T, Medvedev O Gels. 2024; 10(12).

PMID: 39727565 PMC: 11675360. DOI: 10.3390/gels10120807.


The reliability of virtual non-contrast reconstructions of photon-counting detector CT scans in assessing abdominal organs.

Dudas I, Schultz L, Benke M, Szucs A, Kaposi P, Szijarto A BMC Med Imaging. 2024; 24(1):237.

PMID: 39251996 PMC: 11386360. DOI: 10.1186/s12880-024-01419-w.


Exploring the Potential of a Novel Iodine-Based Material as an Alternative Contrast Agent in X-ray Imaging Studies.

Bliznakova K, Kolev I, Dukov N, Dimova T, Bliznakov Z Materials (Basel). 2024; 17(9).

PMID: 38730863 PMC: 11084318. DOI: 10.3390/ma17092059.


Evaluation of hafnium oxide nanoparticles imaging characteristics as a contrast agent in X-ray computed tomography.

Safari A, Mahdavi M, Fardid R, Oveisi A, Jalli R, Haghani M Radiol Phys Technol. 2024; 17(2):441-450.

PMID: 38630390 DOI: 10.1007/s12194-024-00797-8.