» Articles » PMID: 33011206

Identification of Treatment Error Types for Lung Cancer Patients Using Convolutional Neural Networks and EPID Dosimetry

Overview
Journal Radiother Oncol
Specialties Oncology
Radiology
Date 2020 Oct 4
PMID 33011206
Citations 10
Authors
Affiliations
Soon will be listed here.
Abstract

Background/purpose: Electronic portal imaging device (EPID) dosimetry aims to detect treatment errors, potentially leading to treatment adaptation. Clinically used threshold classification methods for detecting errors lead to loss of information (from multi-dimensional EPID data to a few numbers) and cannot be used for identifying causes of errors. Advanced classification methods, such as deep learning, can use all available information. In this study, convolutional neural networks (CNNs) were trained to detect and identify error type and magnitude of simulated treatment errors in lung cancer patients. The purpose of this simulation study is to provide a proof-of-concept of CNNs for error identification using EPID dosimetry in an in vivo scenario.

Materials And Methods: Clinically realistic ranges of anatomical changes, positioning errors and mechanical errors were simulated for lung cancer patients. Predicted portal dose images (PDIs) containing errors were compared to error-free PDIs using the widely used gamma analysis. CNNs were trained to classify errors using 2D gamma maps. Three classification levels were assessed: Level 1 (main error type, e.g., anatomical change), Level 2 (error subtype, e.g., tumor regression) and Level 3 (error magnitude, e.g., >50% tumor regression).

Results: CNNs showed good performance for all classification levels (training/test accuracy 99.5%/96.1%, 92.5%/86.8%, 82.0%/72.9%). For Level 3, overfitting became more apparent.

Conclusion: This simulation study indicates that deep learning is a promising powerful tool for identifying types and magnitude of treatment errors with EPID dosimetry, providing additional information not currently available from EPID dosimetry. This is a first step towards rapid, automated models for identification of treatment errors using EPID dosimetry.

Citing Articles

A Thorough Review of the Clinical Applications of Artificial Intelligence in Lung Cancer.

Kotoulas S, Spyratos D, Porpodis K, Domvri K, Boutou A, Kaimakamis E Cancers (Basel). 2025; 17(5).

PMID: 40075729 PMC: 11898928. DOI: 10.3390/cancers17050882.


"Under the hood": artificial intelligence in personalized radiotherapy.

Gianoli C, De Bernardi E, Parodi K BJR Open. 2024; 6(1):tzae017.

PMID: 39104573 PMC: 11299549. DOI: 10.1093/bjro/tzae017.


Treatment plan complexity quantification for predicting gamma passing rates in patient-specific quality assurance for stereotactic volumetric modulated arc therapy.

Xue X, Luan S, Ding Y, Li X, Li D, Wang J J Appl Clin Med Phys. 2024; 25(9):e14432.

PMID: 38889335 PMC: 11492345. DOI: 10.1002/acm2.14432.


Error detection for radiotherapy planning validation based on deep learning networks.

Liu S, Ma J, Tang F, Liang Y, Li Y, Li Z J Appl Clin Med Phys. 2024; 25(8):e14372.

PMID: 38709158 PMC: 11302817. DOI: 10.1002/acm2.14372.


In vivo EPID-based daily treatment error identification for volumetric-modulated arc therapy in head and neck cancers with a hierarchical convolutional neural network: a feasibility study.

Zeng Y, Li H, Chang Y, Han Y, Liu H, Pang B Phys Eng Sci Med. 2024; 47(3):907-917.

PMID: 38647634 DOI: 10.1007/s13246-024-01414-z.