Refractive Stability of a New Single-piece Hydrophobic Acrylic Intraocular Lens and Corneal Wound Repair After Implantation Using a New Automated Intraocular Lens Delivery System

Overview
Journal PLoS One
Date 2020 Sep 3
PMID 32877454
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To investigate refractive stability and characterize corneal incision repair up to 3 months after implantation of a new hydrophobic acrylic intraocular lens (IOL) with hydroxyethylmethacrylate using a new automated IOL delivery system.

Methods: This prospective case series included 50 eyes of 50 patients undergoing phacoemulsification and implantation of the Clareon® CNA0T0 IOL using the AutonoMe® automated delivery system in the Department of Ophthalmology, Keio University School of Medicine. The clinical data were collected from 46 eyes of 46 patients preoperatively and 1 day, 1 week, and 1 and 3 months postoperatively. Endothelial-side incision gaping, posterior incision retraction, and Descemet's membrane detachment were recorded as present or absent using anterior-segment optical coherence tomography postoperatively.

Results: The uncorrected distance and corrected distance visual acuities improved and stabilized 1 week postoperatively. The anterior chamber depth was stable from 1 week postoperatively. The subjective refraction was stable from 1 day postoperatively. Descemet's membrane detachments and endothelial-side wound gaping were seen in 19 (41.3%) eyes and 34 (73.9%) eyes 1 day postoperatively and decreased gradually. Posterior incision retraction was seen in eight eyes (17.4%) on day 1 and increased to 19 eyes (41.3%) 3 months postoperatively.

Conclusions: The Clareon IOL had excellent refractive stability from day 1 postoperatively. The AutonoMe automated delivery system enables safe IOL implantation through a 2.4-mm corneal incision, although the wound required longer than 1 month to heal postoperatively.

Citing Articles

Capsular Bag Performance of a Novel Hydrophobic Single-Piece Intraocular Lens.

Waser K, Strassmair K, Pomberger L, Khalil H, Laubichler P, Bolz M Ophthalmol Ther. 2024; 14(2):295-308.

PMID: 39666249 PMC: 11754762. DOI: 10.1007/s40123-024-01075-x.


Visual Outcomes and Patient Satisfaction with a Hydrophobic Acrylic Monofocal IOL Delivered Using a Manual System.

Christenbury J, Hall B Clin Ophthalmol. 2024; 18:3485-3491.

PMID: 39624257 PMC: 11609408. DOI: 10.2147/OPTH.S491589.


Clinical Evaluation of a Hydrophobic Intraocular Lens Using a Preloaded Automated Injector in a Korean Population.

Kim H, Seo K, Yoon K, Choi C, Chung T, Hyon J Clin Ophthalmol. 2023; 17:3353-3363.

PMID: 37941777 PMC: 10629398. DOI: 10.2147/OPTH.S421864.


Evaluation of Visual Outcomes and 3-Month Refractive Stability of a New Hydrophobic Acrylic Intraocular Lens.

Blehm C, Hall B Clin Ophthalmol. 2023; 17:1859-1864.

PMID: 37425027 PMC: 10328104. DOI: 10.2147/OPTH.S415400.


Twelve-Months Follow-Up Postmarket Study of a Hydrophobic Intraocular Lens Using a Preloaded Automated Injector in an Indian Population.

Titiyal J, Basak S, Shetty N, Mathur U, Padmanabhan P, Ganesh S Clin Ophthalmol. 2022; 16:4215-4225.

PMID: 36561374 PMC: 9767064. DOI: 10.2147/OPTH.S379054.


References
1.
Bedar M, Kellner U . [Clinical experience with the Clareon® IOL and the AutonoMe® implantation system]. Ophthalmologe. 2020; 117(11):1100-1104. DOI: 10.1007/s00347-020-01075-9. View

2.
Nanavaty M, Kubrak-Kisza M . Evaluation of preloaded intraocular lens injection systems: Ex vivo study. J Cataract Refract Surg. 2017; 43(4):558-563. DOI: 10.1016/j.jcrs.2017.02.019. View

3.
Xia Y, Liu X, Luo L, Zeng Y, Cai X, Zeng M . Early changes in clear cornea incision after phacoemulsification: an anterior segment optical coherence tomography study. Acta Ophthalmol. 2009; 87(7):764-8. DOI: 10.1111/j.1755-3768.2008.01333.x. View

4.
Miyata K, Kataoka Y, Matsunaga J, Honbo M, Minami K . Prospective Comparison of One-Piece and Three-Piece Tecnis Aspheric Intraocular Lenses: 1-year Stability and its Effect on Visual Function. Curr Eye Res. 2014; 40(9):930-5. DOI: 10.3109/02713683.2014.968936. View

5.
De Giacinto C, Porrelli D, Turco G, Pastore M, DAloisio R, Tognetto D . Surface properties of commercially available hydrophobic acrylic intraocular lenses: Comparative study. J Cataract Refract Surg. 2019; 45(9):1330-1334. DOI: 10.1016/j.jcrs.2019.04.011. View