Complex Confounder-corrected R2* Mapping for Liver Iron Quantification with MRI
Overview
Affiliations
Objectives: MRI-based R2* mapping may enable reliable and rapid quantification of liver iron concentration (LIC). However, the performance and reproducibility of R2* across acquisition protocols remain unknown. Therefore, the objective of this work was to evaluate the performance and reproducibility of complex confounder-corrected R2* across acquisition protocols, at both 1.5 T and 3.0 T.
Methods: In this prospective study, 40 patients with suspected iron overload and 10 healthy controls were recruited with IRB approval and informed written consent and imaged at both 1.5 T and 3.0 T. For each subject, acquisitions included four different R2* mapping protocols at each field strength, and an FDA-approved R2-based method performed at 1.5 T as a reference for LIC. R2* maps were reconstructed from the complex data acquisitions including correction for noise effects and fat signal. For each subject, field strength, and R2* acquisition, R2* measurements were performed in each of the nine liver Couinaud segments and the spleen. R2* measurements were compared across protocols and field strength (1.5 T and 3.0 T), and R2* was calibrated to LIC for each acquisition and field strength.
Results: R2* demonstrated high reproducibility across acquisition protocols (p > 0.05 for 96/108 pairwise comparisons across 2 field strengths and 9 liver segments, ICC > 0.91 for each field strength/segment combination) and high predictive ability (AUC > 0.95 for four clinically relevant LIC thresholds). Calibration of R2* to LIC was LIC = - 0.04 + 2.62 × 10 R2* at 1.5 T and LIC = 0.00 + 1.41 × 10 R2* at 3.0 T.
Conclusions: Complex confounder-corrected R2* mapping enables LIC quantification with high reproducibility across acquisition protocols, at both 1.5 T and 3.0 T.
Key Points: • Confounder-corrected R2* of the liver provides reproducible R2* across acquisition protocols, including different spatial resolutions, echo times, and slice orientations, at both 1.5 T and 3.0 T. • For all acquisition protocols, high correlation with R2-based liver iron concentration (LIC) quantification was observed. • The calibration between confounder-corrected R2* and LIC, at both 1.5 T and 3.0 T, is determined in this study.
Cao S, Zhang J, Dong C, Que Y, Gao X, Guo R Eur Radiol. 2025; .
PMID: 39751630 DOI: 10.1007/s00330-024-11334-x.
Zhou X, Jia X, Chen Y, Song B Biomedicines. 2024; 12(11).
PMID: 39595022 PMC: 11592092. DOI: 10.3390/biomedicines12112456.
Zheng G, Wei F, Lu P, Yang G, Li C, Lin C Quant Imaging Med Surg. 2024; 14(6):3901-3913.
PMID: 38846285 PMC: 11151266. DOI: 10.21037/qims-23-1593.
Serum ferritin and the risk of myocardial infarction: A Mendelian randomization study.
Zhou J Medicine (Baltimore). 2024; 103(17):e37952.
PMID: 38669402 PMC: 11049730. DOI: 10.1097/MD.0000000000037952.
Crawford D, Ramm G, Bridle K, Nicoll A, Delatycki M, Olynyk J Hepatol Int. 2023; 17(3):522-541.
PMID: 37067673 PMC: 10224841. DOI: 10.1007/s12072-023-10510-3.