Noninvasive Measurement of Arterial Blood Pressure in Patients with Continuous-flow Left Ventricular Assist Devices: a Systematic Review

Overview
Journal Heart Fail Rev
Date 2020 Jul 23
PMID 32696152
Citations 2
Authors
Affiliations
Soon will be listed here.
Abstract

Accurate mean blood pressure determination is essential to prevent adverse events in patients with continuous-flow left ventricular assist devices (CFLVAD). We sought to evaluate the accuracy of noninvasive methods of blood pressure measurement compared with invasive intra-arterial recordings in patients with CFLVAD. Systematic electronic search was performed on four online databases (PubMed, Scopus, Embase, and Web of Knowledge) for the terms "Blood Pressure" AND ("Heart-Assist Devices" OR "Left ventricular Assist Devices"). Only studies that compared an intra-arterial and noninvasive blood pressure measurement were included. Electronic search of scientific literature identified 5968 articles. After deduplication, screening of titles and abstracts, full-text review, and excluding incorrect populations and comparator, a total of 12 studies with 502 participants were included, of those 402 participants who had intra-arterial blood pressure measurement. Doppler mean arterial blood pressure showed a very high correlation with mean intra-arterial blood pressure (r = 0.97, r = 0.87) in low pulsatility situations. When the pulsatility was not evaluated, the correlation was high moderate (r = 0.63, r = 0.741). In low pulsatility situations, the correlation was moderate to high moderate (r = 0.42 to r = 0.65). Oscillometer automatic blood pressure cuff showed a moderate to very high correlation with intra-arterial mean arterial blood pressure (r = 0.42, r = 0.86) but also could be low in the context of low pulsatility associated with inconsistent success in noninvasive measurement (r = 0.25). Studies correlating intra-arterial with noninvasive techniques were performed in the context of routine clinical care using fluid-filled catheters. The degree of correlation between both methods is at least moderate.

Citing Articles

Common Complications and Cardiopulmonary Resuscitation in Patients with Left Ventricular Assist Devices: A Narrative Review.

Zaloznik Djordjevic J, Ozkan T, Goncz E, Ksela J, Mockel M, Strnad M Medicina (Kaunas). 2023; 59(11).

PMID: 38004030 PMC: 10672734. DOI: 10.3390/medicina59111981.


Validity and success rate of noninvasive mean arterial blood pressure measurements in cf-LVAD patients: A technical review.

Lankheet S, Pieterse M, Rijnhout R, Tuerlings E, Oppelaar A, van Laake L Artif Organs. 2022; 46(12):2361-2370.

PMID: 35920238 DOI: 10.1111/aor.14367.

References
1.
Muntner P, Shimbo D, Carey R, Charleston J, Gaillard T, Misra S . Measurement of Blood Pressure in Humans: A Scientific Statement From the American Heart Association. Hypertension. 2019; 73(5):e35-e66. PMC: 11409525. DOI: 10.1161/HYP.0000000000000087. View