Introduction: An increased risk of early femoral component loosening has been reported using the direct anterior approach (DAA) compared with other common surgical approaches. However, long-term data are scarce. The purpose of this study is: (1) to determine the incidence of early femoral loosening in a high volume, single surgeon's practice utilising the DAA approach; and (2) to examine the effect of stem design and type of coating on aseptic loosening in the early and mid-term postoperative period.

Methods: A retrospective review of 1650 consecutive patients (1800 hips) who underwent total hip arthroplasty (THA) using the DAA between August 2011 and December 2017 was conducted at our institution. 3 types of uncemented femoral stems (Quadra-S, Avenir, TwinSys), with similar design, but different coating, were implanted. Patients were evaluated clinically and radiologically at 4 weeks, 3 months, 1 year, and annually thereafter.

Results: After a mean follow-up of 46.4 months, the total incidence of revision for aseptic loosening was 0.44% ( = 8). All loose stems were Quadra-S, failing to achieve osseointegration, for an overall incidence of 0.96% ( = 0.002). None of the other stems were loose. Radiolucent lines around the proximal stem portion were visible in 75 Quadra-S stems (4.1%) on radiographs taken at 1 year postoperatively and continued to deteriorate at the latest follow-up.

Conclusions: In this specific cohort of patients, the increased rate of femoral stem aseptic loosening was implant-related and was attributed only to a specific type of femoral stem (Quadra-S). No relation to other factors was proven suggesting that the surface characteristics of this femoral stem and the lack of bioactive coating are responsible for the observed early femoral failures. These findings should be confirmed by additional registry work and larger population sample sizes are needed to evaluate the prosthesis performance after implantation through the DAA.

Citing Articles

Impact of femoral stem alignment on periprosthetic bone density in THA: a study of the Avenir Complete stem.

Maeda T, Kuroda Y, Kamenaga T, Matsumoto T, Kuroda R, Hayashi S Arch Orthop Trauma Surg. 2024; 145(1):74.

PMID: 39708142 DOI: 10.1007/s00402-024-05617-7.


Correlation of cross-linking as measured by the trans-vinylene index and in vitro wear of polyethylene from various acetabular liner brands.

Wahl P, Heuberger R, Pascucci A, Imwinkelried T, Furstner M, Icken N Bone Joint Res. 2024; 13(11):682-693.

PMID: 39588944 PMC: 11590334. DOI: 10.1302/2046-3758.1311.BJR-2024-0203.R1.


Monte Carlo-based in-depth morphological analysis of medullary cavity for designing personalized femoral stem.

Wang L, Sun H, Guo K, He K, Geng W, Zhou W Front Surg. 2024; 11:1294749.

PMID: 39183780 PMC: 11341490. DOI: 10.3389/fsurg.2024.1294749.


Changes in serum inflammatory factors after hip arthroplasty and analysis of risk factors for prosthesis loosening.

He M, Zhu Q, Yin D, Duan Y, Sun P, Fang Q Am J Transl Res. 2024; 16(2):557-566.

PMID: 38463599 PMC: 10918134.


Global mapping of institutional and hospital-based (Level II-IV) arthroplasty registries: a scoping review.

Zgouridou A, Kenanidis E, Potoupnis M, Tsiridis E Eur J Orthop Surg Traumatol. 2023; 34(2):1219-1251.

PMID: 37768398 PMC: 10858160. DOI: 10.1007/s00590-023-03691-y.