Short-term Reproducibility of Ambulatory Blood Pressure Measurements: a Systematic Review and Meta-analysis of 35 Observational Studies
Overview
Authors
Affiliations
Objective: A systematic review on the reproducibility of ambulatory blood pressure measurements (ABPM) has not yet been conducted. This meta-analysis compared 24-h/daytime/night-time SBP and DBP mean values and SBP/DBP nocturnal dipping status from ABPMs in participants with or without hypertension.
Methods: Ovid MEDLINE, EMBASE, and CINAHL Complete databases were searched for articles published before 3 May 2019. Eligible studies reporting a 24-h ABPM repeated at least once within 1 month were included. The mean daytime/night-time/24-h BP values, percentage of nocturnal dipping, and proportion of nondippers were compared between the first and second day of measurements, and the proportion of participants with inconsistent dipping status were estimated using a random effect model.
Results: Population-based analysis found a 0-1.1 mmHg difference between the first and second ABPM for 24-h/daytime/night-time SBP and DBP and 0-0.5% for percentage of SBP/DBP nocturnal dipping. The proportion of non-dippers was not different between the first and second ABPM. Intra-individual analysis found that the 95% limit of agreements (LOA) for SBP/DBP were wide and the 95% LOA for daytime SBP, common reference to diagnose hypertension, ranged -16.7 to 18.4 mmHg. Similarly, 32% of participants had inconsistent nocturnal dipping status.
Conclusion: ABPM had excellent reproducibility at the population level, favouring its application for research purposes; but reproducibility of intra-individual BP values and dipping status from a 24-h ABPM was limited. The available evidence was limited by the lack of high-quality studies and lack of studies in non-Western populations.
Dipping status assessed by home vs. ambulatory blood pressure monitoring.
Cheng Y, Li Y Hypertens Res. 2025; .
PMID: 39939828 DOI: 10.1038/s41440-025-02139-6.
Clinical practice guideline for the management of hypertension in China.
Chin Med J (Engl). 2024; 137(24):2907-2952.
PMID: 39653517 PMC: 11706600. DOI: 10.1097/CM9.0000000000003431.
Pilz N, Narkiewicz K, Wolf J, Kario K, Visser T, Opatz O Hypertens Res. 2024; 48(3):1144-1154.
PMID: 39639129 PMC: 11879860. DOI: 10.1038/s41440-024-02056-0.
Palatini P, Battista F, Mos L, Rattazzi M, Ermolao A, Vriz O J Hypertens. 2024; 43(1):128-135.
PMID: 39288253 PMC: 11608614. DOI: 10.1097/HJH.0000000000003874.
Wearable cuffless blood pressure tracking: when will they be good enough?.
Schutte A J Hum Hypertens. 2024; 38(9):669-672.
PMID: 38997475 PMC: 11387187. DOI: 10.1038/s41371-024-00932-3.