» Articles » PMID: 32517226

Estimation of Heart Rate and Respiratory Rate from PPG Signal Using Complementary Ensemble Empirical Mode Decomposition with Both Independent Component Analysis and Non-Negative Matrix Factorization

Overview
Journal Sensors (Basel)
Publisher MDPI
Specialty Biotechnology
Date 2020 Jun 11
PMID 32517226
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

This paper proposes a framework combining the complementary ensemble empirical mode decomposition with both the independent component analysis and the non-negative matrix factorization for estimating both the heart rate and the respiratory rate from the photoplethysmography (PPG) signal. After performing the complementary ensemble empirical mode decomposition on the PPG signal, a finite number of intrinsic mode functions are obtained. Then, these intrinsic mode functions are divided into two groups to perform the further analysis via both the independent component analysis and the non-negative matrix factorization. The surrogate cardiac signal related to the heart activity and another surrogate respiratory signal related to the respiratory activity are reconstructed to estimate the heart rate and the respiratory rate, respectively. Finally, different records of signals acquired from the Medical Information Mart for Intensive Care database downloaded from the Physionet Automated Teller Machine (ATM) data bank are employed for demonstrating the outperformance of our proposed method. The results show that our proposed method outperforms both the digital filtering approach and the conventional empirical mode decomposition based methods in terms of reconstructing both the surrogate cardiac signal and the respiratory signal from the PPG signal as well as both achieving the higher accuracy and the higher reliability for estimating both the heart rate and the respiratory rate.

Citing Articles

CogWatch: An open-source platform to monitor physiological indicators for cognitive workload and stress.

Dankovich 4th L, Joyner J, He W, Sesay A, Vaughn-Cooke M HardwareX. 2024; 19:e00538.

PMID: 38962730 PMC: 11220525. DOI: 10.1016/j.ohx.2024.e00538.


Continuous Critical Respiratory Parameter Measurements Using a Single Low-Cost Relative Humidity Sensor: Evaluation Study.

Vaussenat F, Bhattacharya A, Payette J, Benavides-Guerrero J, Perrotton A, Gerlein L JMIR Biomed Eng. 2024; 8:e47146.

PMID: 38875670 PMC: 11041423. DOI: 10.2196/47146.


Assessing heart rate and blood pressure estimation from image photoplethysmography using a digital blood pressure meter.

Trirongjitmoah S, Promking A, Kaewdang K, Phansiri N, Treeprapin K Heliyon. 2024; 10(5):e27113.

PMID: 38439889 PMC: 10909774. DOI: 10.1016/j.heliyon.2024.e27113.


An Optical Signal Simulator for the Characterization of Photoplethysmographic Devices.

Pittella E, Testa O, Podesta L, Piuzzi E Sensors (Basel). 2024; 24(3).

PMID: 38339729 PMC: 10857427. DOI: 10.3390/s24031008.


Wearable Sensors as a Preoperative Assessment Tool: A Review.

Syversen A, Dosis A, Jayne D, Zhang Z Sensors (Basel). 2024; 24(2).

PMID: 38257579 PMC: 10820534. DOI: 10.3390/s24020482.


References
1.
Orphanidou C . Derivation of respiration rate from ambulatory ECG and PPG using Ensemble Empirical Mode Decomposition: Comparison and fusion. Comput Biol Med. 2016; 81:45-54. DOI: 10.1016/j.compbiomed.2016.12.005. View

2.
Nilsson L, Johansson A, Kalman S . Monitoring of respiratory rate in postoperative care using a new photoplethysmographic technique. J Clin Monit Comput. 2003; 16(4):309-15. DOI: 10.1023/a:1011424732717. View

3.
Nakajima K, Tamura T, Miike H . Monitoring of heart and respiratory rates by photoplethysmography using a digital filtering technique. Med Eng Phys. 1996; 18(5):365-72. DOI: 10.1016/1350-4533(95)00066-6. View

4.
See V, Kwong H . Wearing Your Heart on Your Wrist: How Wearable Smart Devices Are Shaping the Landscape of Early Cardiac Arrhythmia Detection. JACC Case Rep. 2021; 2(3):434-436. PMC: 8311711. DOI: 10.1016/j.jaccas.2019.12.037. View

5.
Leonard P, Grubb N, Addison P, Clifton D, Watson J . An algorithm for the detection of individual breaths from the pulse oximeter waveform. J Clin Monit Comput. 2005; 18(5-6):309-12. DOI: 10.1007/s10877-005-2697-z. View