Pulse Sequences As Tissue Property Filters (TP-filters): a Way of Understanding the Signal, Contrast and Weighting of Magnetic Resonance Images

Overview
Specialty Radiology
Date 2020 Jun 4
PMID 32489930
Citations 9
Authors
Affiliations
Soon will be listed here.
Abstract

This paper describes a quantitative approach to understanding the signal, contrast and weighting of magnetic resonance (MR) images. It uses the concept of pulse sequences as tissue property (TP) filters and models the signal, contrast and weighting of sequences using either a single TP-filter (univariate model) or several TP-filters (the multivariate model). For the spin echo (SE) sequence using the Bloch equations, voxel signal intensity is plotted against the logarithm of the value of the TPs contributing to the sequence signal to produce three TP-filters, an exponential ρ-filter, a low pass T-filter and a high pass T-filter. Using the univariate model which considers signal changes in only one of ρ, T, or T at a time, the first partial derivative of signal with respect to the natural logarithm of ρ, T or T is the sequence weighting for each filter (for small changes in each TP). Absolute contrast is then the sequence weighting multiplied by the fractional change in TP for each filter. For large changes in TPs, the same approach is followed, but using the mean slope of the filter as the sequence weighting. These approaches can also be used for fractional contrast. The univariate TP-filter model provides a mathematical framework for converting conventional qualitative univariate weighting as used in everyday clinical practice into quantitative univariate weighting. Using the multivariate model which considers several TP-filters together, the relative contributions of each TP to overall sequence and image weighting are expressed as sequence and imaging weighting ratios respectively. This is not possible with conventional qualitative weighting which is univariate. The same approaches are used for inversion recovery (IR), pulsed gradient SE, spoiled gradient echo (SGE), balanced steady state free precession, ultrashort echo time and other pulse sequences. Other TPs such as susceptibility, chemical shift and flow can be included with phase along the Y axis of the TP-filter. Contrast agent effects are also included. In the text TP-filters are distinguished from k-space filters, signal filters (S-filters) which are used in imaging processing as well as to describe windowing the signal width and level of images, and spatial filters. The TP-filters approach resolves many of the ambiguities and inconsistencies associated with conventional qualitative weighting and provides a variety of new insights into the signal, contrast and weighting of MR images which are not apparent using qualitative weighting. The TP-filter approach relates the preparation component of pulse sequences to voxel signal, and contrast between two voxels. This is complementary to k-space which relates the acquisition component of pulse sequences to the spatial properties of MR images and their global contrast.

Citing Articles

Yet more evidence that non-aqueous myelin lipids can be directly imaged with ultrashort echo time (UTE) MRI on a clinical 3T scanner: a lyophilized red blood cell membrane lipid study.

Shin S, Moazamian D, Suprana A, Zeng C, Athertya J, Carl M Neuroimage. 2024; 296:120666.

PMID: 38830440 PMC: 11380916. DOI: 10.1016/j.neuroimage.2024.120666.


Ultra-High Contrast MRI: Using Divided Subtracted Inversion Recovery (dSIR) and Divided Echo Subtraction (dES) Sequences to Study the Brain and Musculoskeletal System.

Cornfeld D, Condron P, Newburn G, McGeown J, Scadeng M, Bydder M Bioengineering (Basel). 2024; 11(5).

PMID: 38790308 PMC: 11118255. DOI: 10.3390/bioengineering11050441.


Targeted magnetic resonance imaging (tMRI) of small changes in the T and spatial properties of normal or near normal appearing white and gray matter in disease of the brain using divided subtracted inversion recovery (dSIR) and divided reverse....

Ma Y, Moazamian D, Port J, Edjlali M, Pruvo J, Hacein-Bey L Quant Imaging Med Surg. 2023; 13(10):7304-7337.

PMID: 37869282 PMC: 10585510. DOI: 10.21037/qims-23-232.


Making the invisible visible-ultrashort echo time magnetic resonance imaging: Technical developments and applications.

Ma Y, Jang H, Jerban S, Chang E, Chung C, Bydder G Appl Phys Rev. 2022; 9(4):041303.

PMID: 36467869 PMC: 9677812. DOI: 10.1063/5.0086459.


Improving the understanding and performance of clinical MRI using tissue property filters and the central contrast theorem, MASDIR pulse sequences and synergistic contrast MRI.

Ma Y, Moazamian D, Cornfeld D, Condron P, Holdsworth S, Bydder M Quant Imaging Med Surg. 2022; 12(9):4658-4690.

PMID: 36060593 PMC: 9403590. DOI: 10.21037/qims-22-394.


References
1.
Twieg D . The k-trajectory formulation of the NMR imaging process with applications in analysis and synthesis of imaging methods. Med Phys. 1983; 10(5):610-21. DOI: 10.1118/1.595331. View

2.
Axel L . Revised glossary of MR terms. Radiology. 1987; 162(3):874. DOI: 10.1148/radiology.162.3.3809508. View

3.
Reichert I, Robson M, Gatehouse P, He T, Chappell K, Holmes J . Magnetic resonance imaging of cortical bone with ultrashort TE pulse sequences. Magn Reson Imaging. 2005; 23(5):611-8. DOI: 10.1016/j.mri.2005.02.017. View

4.
Ma Y, Searleman A, Jang H, Wong J, Chang E, Corey-Bloom J . Whole-Brain Myelin Imaging Using 3D Double-Echo Sliding Inversion Recovery Ultrashort Echo Time (DESIRE UTE) MRI. Radiology. 2019; 294(2):362-374. PMC: 6996715. DOI: 10.1148/radiol.2019190911. View

5.
Rooney W, Johnson G, Li X, Cohen E, Kim S, Ugurbil K . Magnetic field and tissue dependencies of human brain longitudinal 1H2O relaxation in vivo. Magn Reson Med. 2007; 57(2):308-18. DOI: 10.1002/mrm.21122. View