» Articles » PMID: 32435467

Serum Cobalt Concentrations Remain at Low Levels at a Minimum of 20 Years Following Metal-on-metal Total Hip Arthroplasty

Overview
Journal Bone Joint Res
Date 2020 May 22
PMID 32435467
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

Aims: Second-generation metal-on-metal (MoM) articulations in total hip arthroplasty (THA) were introduced in order to reduce wear-related complications. The current study reports on the serum cobalt levels and the clinical outcome at a minimum of 20 years following THA with a MoM (Metasul) or a ceramic-on-polyethylene (CoP) bearing.

Methods: The present study provides an update of a previously published prospective randomized controlled study, evaluating the serum cobalt levels of a consecutive cohort of 100 patients following THA with a MoM or a CoP articulation. A total of 31 patients were available for clinical and radiological follow-up examination. After exclusion of 11 patients because of other cobalt-containing implants, 20 patients (MoM (n = 11); CoP (n = 9)) with a mean age of 69 years (42 to 97) were analyzed. Serum cobalt levels were compared to serum cobalt levels five years out of surgery.

Results: The median cobalt concentration in the MoM group was 1.04 μg/l (interquartile range (IQR) 0.64 to 1.70) at a mean of 21 years (20 to 24) postoperatively and these values were similar (p = 0.799) to cobalt levels at five years. In the CoP control group, the median cobalt levels were below the detection limit (< 0.3 μg/l; median 0.15 μg/l, IQR 0.15 to 0.75) at 20 years. The mean Harris Hip Score was 91.4 points (61 to 100) in the MoM group and 92.8 points (63 to 100) in the CoP group.

Conclusion: This study represents the longest follow-up series evaluating the serum cobalt levels after 28 mm head MoM bearing THA and shows that serum cobalt concentrations remain at low levels at a mean of 21 years (20 to 24) after implantation. 2020;9(3):145-150.

Citing Articles

Adverse Tissue Reactions and Metal Ion Behavior After Small-Head Metasul Hip Arthroplasty: A Long-Term Follow-Up Study.

Ishida T, Tateiwa T, Takahashi Y, Nishikawa Y, Shishido T, Masaoka T Orthop Surg. 2021; 14(2):264-273.

PMID: 34910382 PMC: 8867442. DOI: 10.1111/os.13125.


Course of Metal Ions after a Revision of Malfunctioning Metal-on-Metal Total Hip Prostheses.

Nicolli A, Bortoletti I, Maso S, Trevisan A Medicina (Kaunas). 2021; 57(2).

PMID: 33525335 PMC: 7912175. DOI: 10.3390/medicina57020115.


Favourable clinical outcomes following cemented arthroplasty after metal-on-metal total hip replacement: a retrospective study with a mean follow-up of 10 years.

Yu W, Chen M, Zeng X, Zhao M, Zhang X, Ye J BMC Musculoskelet Disord. 2020; 21(1):772.

PMID: 33220707 PMC: 7680591. DOI: 10.1186/s12891-020-03797-y.

References
1.
Engh C, Bobyn J, Glassman A . Porous-coated hip replacement. The factors governing bone ingrowth, stress shielding, and clinical results. J Bone Joint Surg Br. 1987; 69(1):45-55. DOI: 10.1302/0301-620X.69B1.3818732. View

2.
Barlow B, Ortiz P, Boles J, Lee Y, Padgett D, Westrich G . What Are Normal Metal Ion Levels After Total Hip Arthroplasty? A Serologic Analysis of Four Bearing Surfaces. J Arthroplasty. 2017; 32(5):1535-1542. DOI: 10.1016/j.arth.2016.11.024. View

3.
Nam D, Keeney J, Nunley R, Johnson S, Clohisy J, Barrack R . Metal Ion Concentrations in Young, Active Patients Following Total Hip Arthroplasty with the Use of Modern Bearing Couples. J Arthroplasty. 2015; 30(12):2227-32. DOI: 10.1016/j.arth.2015.06.025. View

4.
Randelli F, Banci L, DAnna A, Visentin O, Randelli G . Cementless Metasul metal-on-metal total hip arthroplasties at 13 years. J Arthroplasty. 2011; 27(2):186-92. DOI: 10.1016/j.arth.2011.04.015. View

5.
Langton D, Jameson S, Joyce T, Hallab N, Natu S, Nargol A . Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: A consequence of excess wear. J Bone Joint Surg Br. 2010; 92(1):38-46. DOI: 10.1302/0301-620X.92B1.22770. View