On the Origin and Correction for Inner Filter Effects in Fluorescence Part I: Primary Inner Filter Effect-the Proper Approach for Sample Absorbance Correction

Overview
Publisher IOP Publishing
Specialties Biochemistry
Chemistry
Date 2020 May 20
PMID 32428893
Citations 12
Authors
Affiliations
Soon will be listed here.
Abstract

Fluorescence technologies have been the preferred method for detection, analytical sensing, medical diagnostics, biotechnology, imaging, and gene expression for many years. Fluorescence becomes essential for studying molecular processes with high specificity and sensitivity through a variety of biological processes. A significant problem for practical fluorescence applications is the apparent non-linearity of the fluorescence intensity resulting from inner-filter effects, sample scattering, and absorption of intrinsic components of biological samples. Sample absorption can lead to the primary inner filter effect (Type I inner filter effect) and is the first factor that should be considered. This is a relatively simple factor to be controlled in any fluorescence experiment. However, many previous approaches have given only approximate experimental methods for correcting the deviation from expected results. In this part we are discussing the origin of the primary inner filter effect and presenting a universal approach for correcting the fluorescence intensity signal in the full absorption range. Importantly, we present direct experimental results of how the correction works. One considers problems emerging from varying absorption across its absorption spectrum for all fluorophores. We use Rhodamine 800 and demonstrate how to properly correct the excitation spectra in a broad wavelength range. Second is the effect of an inert absorber that attenuates the intensity of the excitation beam as it travels through the cuvette, which leads to a significant deviation of observed results. As an example, we are presenting fluorescence quenching of a tryptophan analog, NATA, by acrylamide and we show how properly corrected results compare to the initial erroneous results. The procedure is generic and applies to many other applications like quantum yield determination, tissue/blood absorption, or acceptor absorption in FRET experiments.

Citing Articles

Interaction between pristine nC and bovine serum albumin by fluorimetry: assessment of inner filter effect corrections.

Zhang Y, Zheng Y, Li Y, Liu S Front Bioeng Biotechnol. 2025; 13:1518698.

PMID: 40051838 PMC: 11882559. DOI: 10.3389/fbioe.2025.1518698.


Quantifying glucose uptake at the single cell level with confocal microscopy reveals significant variability within and across individuals.

Paprocki J, Macdonald P, Xu Y, Cheng A, Dunn T, Tetin S Sci Rep. 2025; 15(1):2661.

PMID: 39837851 PMC: 11751187. DOI: 10.1038/s41598-024-74574-3.


Assessment of optimized FRET substrates as universal corona- and picornavirus main protease substrates for screening assays.

Fischer C, Van Oers T, van Belkum M, Lamer T, Romney A, Chen P RSC Adv. 2024; 14(48):35438-35446.

PMID: 39502179 PMC: 11537199. DOI: 10.1039/d4ra06573e.


Development of an advanced multiwavelength emission detector for the analytical ultracentrifuge.

Lautenbach V, Onishchukov G, Wawra S, Frank U, Hartmann L, Peukert W Nanoscale Adv. 2024; 6(10):2611-2622.

PMID: 38752146 PMC: 11093262. DOI: 10.1039/d3na00980g.


Reducing the Inner Filter Effect in Microplates by Increasing Absorbance? Linear Fluorescence in Highly Concentrated Fluorophore Solutions in the Presence of an Added Absorber.

Friganovic T, Weitner T Anal Chem. 2023; 95(35):13036-13045.

PMID: 37606558 PMC: 10483458. DOI: 10.1021/acs.analchem.3c01295.