Reactions of Surrounding Tissue to the Cementless Hip Implant Ti-6Al-4V After an Implantation Period of Several Years. Autopsy Studies in Three Cases

Overview
Specialty Orthopedics
Date 1988 Jan 1
PMID 3240076
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

Three femoral endoprosthetic stems implanted without cement, made of Ti-6Al-4V alloy, with implantation periods of between 2 and 4 years were examined histologically in the form of undecalcified thin ground sections after having been split into segments and embedded in plastic. During the implantation period total osseointegration in the metadiaphyseal region takes place, while in the proximal area reaction forms with direct contact between metal and bone on the one hand and with an interposed membrane of connective tissue and adjacent secondary bony shell on the other must be distinguished. The osseointegration of the stem is the result of a primary healing of the bone and is due to the biocompatibility of the metal alloy and to the primary stability obtained by the press-fit method.

Citing Articles

Evaluation of peri-prosthetic radiolucent lines surrounding the cementless femoral stem using digital tomosynthesis with metal artifact reduction: a cadaveric study in comparison with radiography and computed tomography.

Tang H, Huang X, Cheng X, Yang D, Huang Y, Zhou Y Quant Imaging Med Surg. 2020; 10(9):1786-1800.

PMID: 32879857 PMC: 7417763. DOI: 10.21037/qims-19-1018.


Late removal of titanium hardware from the elbow is problematic.

Bachoura A, Yoshida R, Lattermann C, Kamineni S ISRN Orthop. 2014; 2012:256239.

PMID: 24977074 PMC: 4063128. DOI: 10.5402/2012/256239.


Effect of the initial implant fitting on the predicted secondary stability of a cementless stem.

Viceconti M, Pancanti A, Dotti M, Traina F, Cristofolini L Med Biol Eng Comput. 2004; 42(2):222-9.

PMID: 15125153 DOI: 10.1007/BF02344635.

References
1.
Albrektsson T, Branemark P, Hansson H, Lindstrom J . Osseointegrated titanium implants. Requirements for ensuring a long-lasting, direct bone-to-implant anchorage in man. Acta Orthop Scand. 1981; 52(2):155-70. DOI: 10.3109/17453678108991776. View

2.
Zweymuller K, Semlitsch M . Concept and material properties of a cementless hip prosthesis system with Al2O3 ceramic ball heads and wrought Ti-6Al-4V stems. Arch Orthop Trauma Surg (1978). 1982; 100(4):229-36. DOI: 10.1007/BF00381662. View

3.
Lintner F, Zweymuller K, Brand G . Tissue reactions to titanium endoprostheses. Autopsy studies in four cases. J Arthroplasty. 1986; 1(3):183-95. DOI: 10.1016/s0883-5403(86)80029-8. View

4.
Aldinger G, Mitzkat K . [The effect of aging on endoprostheses]. Z Orthop Ihre Grenzgeb. 1986; 124(4):392-5. View

5.
Zweymuller K . A cementless titanium hip endoprosthesis system based on press-fit fixation: basic research and clinical results. Instr Course Lect. 1986; 35:203-25. View